Stability over time of immature platelet fraction and comparison between EDTA and citrated whole blood samples.
Noel, Marc Michel; Feugray, Guillaume; Kasonga, Fiston; Buchonnet, Gérard; Daliphard, Sylvie; Barbay, Virginie; Bera, Elsa; Le Cam Duchez, Véronique; Billoir, Paul.
Afiliación
  • Noel MM; CHU Rouen, Vascular Hemostasis Unit, Rouen, France.
  • Feugray G; Univ ROUEN, INSERM U1096, Rouen, France.
  • Kasonga F; General Biochemistry, CHU Rouen, Rouen, France.
  • Buchonnet G; CHU Rouen, Vascular Hemostasis Unit, Rouen, France.
  • Daliphard S; Rouen, France Hematology laboratory, CHU Rouen, Rouen, France.
  • Barbay V; Rouen, France Hematology laboratory, CHU Rouen, Rouen, France.
  • Bera E; CHU Rouen, Vascular Hemostasis Unit, Rouen, France.
  • Le Cam Duchez V; Rouen, France Hematology laboratory, CHU Rouen, Rouen, France.
  • Billoir P; CHU Rouen, Vascular Hemostasis Unit, Rouen, France.
J Clin Lab Anal ; 37(13-14): e24946, 2023 Jul.
Article en En | MEDLINE | ID: mdl-37475131
BACKGROUND: Immature platelets (IP) are the youngest circulating platelets, released from megakaryocytes, and demonstrating increased dimensions, significant RNA content, and enhanced activity. Immature platelet research focuses on a differential diagnostic help in patients with thrombocytopenia. The objectives of this study were to compare the variability of IP in citrate and EDTA samples, and to determine stability over time. METHODS: Fifty-six patients were included for comparison between EDTA and citrate whole blood sample collection. Among the patients, 28 had thrombocytopenia (platelet count < 150G/L). Platelet measurement impedancemetry and fluorimetry were performed with Sysmex XN-9000. The immature platelet fraction (IPF) and absolute immature platelet count (A-IPC) were determined with a fluorescent method. RESULTS: The mean value of platelet count with fluorescence was, in EDTA sample, 215 ± 171 and, in citrate sample, 153 ± 118 G/L. No significant difference was observed between IPF between EDTA and citrate (7.74 ± 6.68% vs. 8.45 ± 7.37%, p = 0.69), respectively. With the Bland-Altman analysis, the mean difference in the EDTA sample, between 1 and 24 h, was 8.06 ± 6.96% and 8.73 ± 7.12% for IPF, whereas in the citrate sample, between 1 and 6 h, it was 8.60 ± 7.29% and 7.54 ± 6.97%, for IPF. Comparing 1 h EDTA sample with 6 h citrate sample, the variance ratio was 0.974 (95% CI: 0.864-1.084) in IPF. CONCLUSIONS: We confirmed the potential to conduct IP measurements up to 24 h in the EDTA sample and IPF measurements in the citrate sample for up to 6 h. These results may be useful for the use of IPF, which is a promising parameter whose interest in clinical practice and standardization is not yet well defined.