Gaussian optics optimization of an optical particle counter for measurement in high-pressure gas.
Appl Opt ; 62(17): 4624-4634, 2023 Jun 10.
Article en En | MEDLINE | ID: mdl-37707160
ABSTRACT
An optimization model for the parameters of the pressure-resistant lens of an optical particle counter, based on Gaussian optics theory, was established to increase the measurement accuracy of the counter for high-pressure natural gas. Comparing the experimental and calculated values of the calibrated model, when the pressure-resistant lens is displaced by 2 mm under atmospheric pressure, the relative error of the measured body deformation is 0.15%. When the air pressure varies in the range 0.10-5.09 MPa, the maximum relative error of optical measurement volume deformation with the change of refractive index is 0.13%, which shows that the model has high reliability and accuracy.

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Idioma: En Revista: Appl Opt Año: 2023 Tipo del documento: Article