Five-year outcomes for bell bottom, iliac branch endoprosthesis, and coil and cover approaches from the GREAT registry.
J Vasc Surg
; 79(6): 1369-1378, 2024 Jun.
Article
en En
| MEDLINE
| ID: mdl-38316346
ABSTRACT
OBJECTIVE:
There are a variety of methods used today to treat common iliac aneurysms with endovascular techniques. Of these approaches, little is known about whether a particular limb strategy influences endoleak, reintervention, or aneurysm regression rates. We present 5-year data comparing endoleak, stent graft migration, fracture, aneurysm sac dynamics, and aortic rupture rates among patients treated with bell bottom limbs (BB), iliac branch endoprosthesis (IBE), and coil and cover (CC) approaches from the Global Registry for Endovascular Aortic Treatment registry. Secondary end points were all-cause mortality, stroke, reintervention, and paraplegia.METHODS:
Subjects from the GORE Global Registry for Endovascular Aortic Treatment were enrolled over a 5-year period from October 2017 to August 2022. We included 924 subjects in this study. Statistical data was generated on R software and limb groups were compared using the Pearson's χ2 test and the Kruskal-Wallis rank-sum test.RESULTS:
We found no statistical difference in endoleak rates, stent graft migration, fracture, or aortic rupture when stratified by limb strategy. There was no difference between limb approaches with regard to aneurysm sac dynamics among those with abdominal aortic aneurysms and common iliac aneurysms. Similarly, no statistical difference between limb strategies was found in all-cause mortality, stroke, paraplegia, or reintervention rates. Among patients that required an additional graft during reintervention, the highest rates were found within the IBE group 8.6%, compared with BB group 2.2% and CC group 1.3% (P = .006).CONCLUSIONS:
Overall, there was no difference among limb strategies in endoleak rates, stent graft migration, aneurysm sac dynamics, aortic rupture rates, or our secondary end points. Increased rates of reintervention requiring an additional graft within the IBE group is noteworthy and must be weighed against the adverse effects of hypogastric sacrifice with the CC approach or potentially less advantageous seal zones in the BB approach. This finding suggests that all limb approaches have equivalent effectiveness in managing the aneurysmal common iliac artery; thus, the choice of limb strategy should be individualized and remain at physician discretion. Future research should include a more robust sample size to reproduce these findings.Palabras clave
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Diseño de Prótesis
/
Prótesis Vascular
/
Stents
/
Sistema de Registros
/
Aneurisma Ilíaco
/
Implantación de Prótesis Vascular
/
Procedimientos Endovasculares
Límite:
Aged
/
Aged80
/
Female
/
Humans
/
Male
Idioma:
En
Revista:
J Vasc Surg
Asunto de la revista:
ANGIOLOGIA
Año:
2024
Tipo del documento:
Article