Implementation and validation of 2-D-array based tests in routine linac quality assurance.
Valenta, Jirí; Schmidt, Manfred; Bert, Christoph.
Afiliación
  • Valenta J; Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Universitätsstraße 27, 91054 Erlangen, Germany; Comprehensive Cancer Center Erlangen-EMN (CCC ER-EMN), Östliche Stadtmauerstraße 30, 91054 Erlangen, Germany. Electronic address: jiri.valenta@fau.de.
  • Schmidt M; Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Universitätsstraße 27, 91054 Erlangen, Germany; Comprehensive Cancer Center Erlangen-EMN (CCC ER-EMN), Östliche Stadtmauerstraße 30, 91054 Erlangen, Germany. Electronic address: manfred.schmidt@uk-erlangen.de.
  • Bert C; Department of Radiation Oncology, Universitätsklinikum Erlangen, Friedrich-Alexander-Universität Erlangen-Nürnberg (FAU), Universitätsstraße 27, 91054 Erlangen, Germany; Comprehensive Cancer Center Erlangen-EMN (CCC ER-EMN), Östliche Stadtmauerstraße 30, 91054 Erlangen, Germany. Electronic address: christoph.bert@uk-erlangen.de.
Z Med Phys ; 2024 Feb 16.
Article en En | MEDLINE | ID: mdl-38368240
ABSTRACT

PURPOSE:

In medical linac quality assurance (QA), to replace film dosimetry with low-resolution 2-D ionization chamber array measurements, to validate the procedures, and to perform a comprehensive sensitivity analysis.

METHODS:

A 2-D ionization chamber array with a spatial resolution of 7.62 mm was deployed to perform the following tests Junction tests, MLC transmission test, beam profile constancy vs. gantry angle test, beam profile constancy vs. low dose delivery test, and beam energy constancy vs. low dose delivery test. Test validation and sensitivity analyses based on short- and long-term statistics of the test results were performed.

RESULTS:

All selected mechanical and dosimetry tests could be successfully performed with a 2-D array. Considering the tolerance limits recommended by the AAPM Task Group 142 report (2009), sensitivities of 99.0% or better and specificities ranging from 99.5% to 99.9% could be achieved in all tests when the proper metrics were chosen.

CONCLUSIONS:

The results showed that a low-resolution 2-D ionization chamber array could replace film dosimetry without having to sacrifice high test sensitivity. Its implementation in the routine clinical linac QA program may involve considerable QA time savings.