Implantable cardioverter-defibrillator lead failure and revision following transcutaneous bicaval valve (TricValve®) implantation.
Plant, Allan; Stewart, Fergus; Hooks, Darren.
Afiliación
  • Plant A; Department of Cardiology, Wellington Regional Hospital, Te Whatu Ora Capital and Coast, Newtown, Wellington, New Zealand.
  • Stewart F; Department of Cardiology, Wellington Regional Hospital, Te Whatu Ora Capital and Coast, Newtown, Wellington, New Zealand.
  • Hooks D; Department of Cardiology, Wellington Regional Hospital, Te Whatu Ora Capital and Coast, Newtown, Wellington, New Zealand.
J Cardiovasc Electrophysiol ; 35(5): 1050-1054, 2024 May.
Article en En | MEDLINE | ID: mdl-38501328
ABSTRACT

INTRODUCTION:

Tricuspid regurgitation is associated with significant morbidity and mortality, and occurs at a higher rate in patients with cardiovascular implantable electronic devices. Percutaneous strategies for managing tricuspid regurgitation are evolving, including the development of bicaval valve implantation which has been successfully used in patients with pacing leads. METHODS AND

RESULTS:

We present the first documented case of lead failure following TricValve® implantation, a dedicated self-expanding system for bicaval valve implantation, and the first successful lead revision procedure in this setting.

CONCLUSION:

The case illustrates important considerations in undertaking percutaneous intervention in patients with cardiovascular implantable electronic devices, and their ongoing management.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Insuficiencia de la Válvula Tricúspide / Desfibriladores Implantables / Falla de Equipo Límite: Aged / Female / Humans / Male Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Nueva Zelanda

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Insuficiencia de la Válvula Tricúspide / Desfibriladores Implantables / Falla de Equipo Límite: Aged / Female / Humans / Male Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Nueva Zelanda