Validation of daily 0.35 T diffusion-weighted MRI for MRI-guided glioblastoma radiotherapy.
Lutsik, Natalia; Nejad-Davarani, Siamak P; Valderrama, Alessandro; Herr, Janette; Maziero, Danilo; Cullison, Kaylie; Azzam, Gregory A; Kubicek, Gregory J; Meshman, Jessica; de la Fuente, Macarena I; Armstrong, Tess; Mellon, Eric A.
Afiliación
  • Lutsik N; Department of Radiation Oncology, Sylvester Comprehensive Cancer Center, University of Miami Miller School of Medicine, Miami, Florida, USA.
  • Nejad-Davarani SP; Department of Radiation Oncology, Sylvester Comprehensive Cancer Center, University of Miami Miller School of Medicine, Miami, Florida, USA.
  • Valderrama A; Department of Radiation Oncology, Sylvester Comprehensive Cancer Center, University of Miami Miller School of Medicine, Miami, Florida, USA.
  • Herr J; Department of Radiation Oncology, Sylvester Comprehensive Cancer Center, University of Miami Miller School of Medicine, Miami, Florida, USA.
  • Maziero D; Department of Radiation Oncology, Sylvester Comprehensive Cancer Center, University of Miami Miller School of Medicine, Miami, Florida, USA.
  • Cullison K; Department of Radiation Medicine & Applied Sciences, UC San Diego Health, La Jolla, California, USA.
  • Azzam GA; Department of Radiation Oncology, Sylvester Comprehensive Cancer Center, University of Miami Miller School of Medicine, Miami, Florida, USA.
  • Kubicek GJ; Department of Radiation Oncology, Sylvester Comprehensive Cancer Center, University of Miami Miller School of Medicine, Miami, Florida, USA.
  • Meshman J; Department of Radiation Oncology, Sylvester Comprehensive Cancer Center, University of Miami Miller School of Medicine, Miami, Florida, USA.
  • de la Fuente MI; Department of Radiation Oncology, Sylvester Comprehensive Cancer Center, University of Miami Miller School of Medicine, Miami, Florida, USA.
  • Armstrong T; Neuro-Oncology division, University of Miami Miller School of Medicine, Miami, Florida, USA.
  • Mellon EA; Former ViewRay, Inc., Ohio, USA.
Med Phys ; 51(8): 5386-5398, 2024 Aug.
Article en En | MEDLINE | ID: mdl-38588475
ABSTRACT

BACKGROUND:

MRI-Linac systems enable daily diffusion-weighed imaging (DWI) MRI scans for assessing glioblastoma tumor changes with radiotherapy treatment.

PURPOSE:

Our study assessed the image quality of echoplanar imaging (EPI)-DWI scans compared with turbo spin echo (TSE)-DWI scans at 0.35 Tesla (T) and compared the apparent diffusion coefficient (ADC) values and distortion of EPI-DWI on 0.35 T MRI-Linac compared to high-field diagnostic MRI scanners.

METHODS:

The calibrated National Institute of Standards and Technology (NIST)/Quantitative Imaging Biomarkers Alliance (QIBA) Diffusion Phantom was scanned on a 0.35 T MRI-Linac, and 1.5 T and 3 T MRI with EPI-DWI. Five patients were scanned on a 0.35 T MRI-Linac with a TSE-DWI sequence, and five other patients were scanned with EPI-DWI on a 0.35 T MRI-Linac and a 3 T MRI. The quality of images was compared between the TSE-DWI and EPI-DWI on the 0.35 T MRI-Linac assessing signal-to-noise ratios and presence of artifacts. EPI-DWI ADC values and distortion magnitude were measured and compared between 0.35 T MRI-Linac and high-field MRI for both phantom and patient studies.

RESULTS:

The average ADC differences between EPI-DWI acquired on the 0.35 T MRI-Linac, 1.5 T and 3 T MRI scanners and published references in the phantom study were 1.7%, 0.4% and 1.0%, respectively. Comparing the ADC values based on EPI-DWI in glioblastoma tumors, there was a 3.36% difference between 0.35 and 3 T measurements. Susceptibility-induced distortions in the EPI-DWI phantoms were 0.46 ± 1.51 mm for 0.35 MRI-Linac, 0.98 ± 0.51 mm for 1.5 T MRI and 1.14 ± 1.88 mm for 3 T MRI; for patients -0.47 ± 0.78 mm for 0.35 T and 1.73 ± 2.11 mm for 3 T MRIs. The mean deformable registration distortion for a phantom was 1.1 ± 0.22 mm, 3.5 ± 0.39 mm and 4.7 ± 0.37 mm for the 0.35 T MRI-Linac, 1.5 T MRI, and 3 T MRI scanners, respectively; for patients this distortion was -0.46 ± 0.57 mm for 0.35 T and 4.2 ± 0.41 mm for 3 T. EPI-DWI 0.35 T MRI-Linac images showed higher SNR and lack of artifacts compared with TSE-DWI, especially at higher b-values up to 1000 s/mm2.

CONCLUSION:

EPI-DWI on a 0.35 T MRI-Linac showed superior image quality compared with TSE-DWI, minor and less distortions than high-field diagnostic scanners, and comparable ADC values in phantoms and glioblastoma tumors. EPI-DWI should be investigated on the 0.35 T MRI-Linac for prediction of early response in patients with glioblastoma.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Glioblastoma / Fantasmas de Imagen / Imagen de Difusión por Resonancia Magnética / Radioterapia Guiada por Imagen Límite: Humans Idioma: En Revista: Med Phys Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Glioblastoma / Fantasmas de Imagen / Imagen de Difusión por Resonancia Magnética / Radioterapia Guiada por Imagen Límite: Humans Idioma: En Revista: Med Phys Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos