Validation of daily 0.35 T diffusion-weighted MRI for MRI-guided glioblastoma radiotherapy.
Med Phys
; 51(8): 5386-5398, 2024 Aug.
Article
en En
| MEDLINE
| ID: mdl-38588475
ABSTRACT
BACKGROUND:
MRI-Linac systems enable daily diffusion-weighed imaging (DWI) MRI scans for assessing glioblastoma tumor changes with radiotherapy treatment.PURPOSE:
Our study assessed the image quality of echoplanar imaging (EPI)-DWI scans compared with turbo spin echo (TSE)-DWI scans at 0.35 Tesla (T) and compared the apparent diffusion coefficient (ADC) values and distortion of EPI-DWI on 0.35 T MRI-Linac compared to high-field diagnostic MRI scanners.METHODS:
The calibrated National Institute of Standards and Technology (NIST)/Quantitative Imaging Biomarkers Alliance (QIBA) Diffusion Phantom was scanned on a 0.35 T MRI-Linac, and 1.5 T and 3 T MRI with EPI-DWI. Five patients were scanned on a 0.35 T MRI-Linac with a TSE-DWI sequence, and five other patients were scanned with EPI-DWI on a 0.35 T MRI-Linac and a 3 T MRI. The quality of images was compared between the TSE-DWI and EPI-DWI on the 0.35 T MRI-Linac assessing signal-to-noise ratios and presence of artifacts. EPI-DWI ADC values and distortion magnitude were measured and compared between 0.35 T MRI-Linac and high-field MRI for both phantom and patient studies.RESULTS:
The average ADC differences between EPI-DWI acquired on the 0.35 T MRI-Linac, 1.5 T and 3 T MRI scanners and published references in the phantom study were 1.7%, 0.4% and 1.0%, respectively. Comparing the ADC values based on EPI-DWI in glioblastoma tumors, there was a 3.36% difference between 0.35 and 3 T measurements. Susceptibility-induced distortions in the EPI-DWI phantoms were 0.46 ± 1.51 mm for 0.35 MRI-Linac, 0.98 ± 0.51 mm for 1.5 T MRI and 1.14 ± 1.88 mm for 3 T MRI; for patients -0.47 ± 0.78 mm for 0.35 T and 1.73 ± 2.11 mm for 3 T MRIs. The mean deformable registration distortion for a phantom was 1.1 ± 0.22 mm, 3.5 ± 0.39 mm and 4.7 ± 0.37 mm for the 0.35 T MRI-Linac, 1.5 T MRI, and 3 T MRI scanners, respectively; for patients this distortion was -0.46 ± 0.57 mm for 0.35 T and 4.2 ± 0.41 mm for 3 T. EPI-DWI 0.35 T MRI-Linac images showed higher SNR and lack of artifacts compared with TSE-DWI, especially at higher b-values up to 1000 s/mm2.CONCLUSION:
EPI-DWI on a 0.35 T MRI-Linac showed superior image quality compared with TSE-DWI, minor and less distortions than high-field diagnostic scanners, and comparable ADC values in phantoms and glioblastoma tumors. EPI-DWI should be investigated on the 0.35 T MRI-Linac for prediction of early response in patients with glioblastoma.Palabras clave
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Glioblastoma
/
Fantasmas de Imagen
/
Imagen de Difusión por Resonancia Magnética
/
Radioterapia Guiada por Imagen
Límite:
Humans
Idioma:
En
Revista:
Med Phys
Año:
2024
Tipo del documento:
Article
País de afiliación:
Estados Unidos