In situ anchored ternary hierarchical hybrid nickel@cobaltous sulfide on poly(3,4-ethylenedioxythiophene)-reduced graphene oxide for highly efficient non-enzymatic glucose sensing.
Sheng, Liying; Tan, Hongtao; Zhu, Licheng; Liu, Kexin; Meng, Alan; Li, Zhenjiang.
Afiliación
  • Sheng L; College of Chemical Engineering, State Key Laboratory Base of Eco-Chemical Engineering, Qingdao University of Science and Technology, Shandong Province, Qingdao, 266042, China.
  • Tan H; College of Materials Science and Engineering, Qingdao University of Science and Technology, Shandong Province, Qingdao, 266042, China.
  • Zhu L; College of Chemical Engineering, State Key Laboratory Base of Eco-Chemical Engineering, Qingdao University of Science and Technology, Shandong Province, Qingdao, 266042, China.
  • Liu K; College of Chemical Engineering, State Key Laboratory Base of Eco-Chemical Engineering, Qingdao University of Science and Technology, Shandong Province, Qingdao, 266042, China.
  • Meng A; College of Chemistry and Molecular Engineering, Key Laboratory of Optic-Electric Sensing and Analytical Chemistry for Life Science, MOE, State Key Laboratory Base of Eco-Chemical Engineering, Qingdao University of Science and Technology, Shandong Province, Qingdao, 266042, China. alanmengqust@163.co
  • Li Z; College of Materials Science and Engineering, Qingdao University of Science and Technology, Shandong Province, Qingdao, 266042, China. zhenjiangli@qust.edu.cn.
Mikrochim Acta ; 191(5): 267, 2024 04 16.
Article en En | MEDLINE | ID: mdl-38627300
ABSTRACT
A ternary hierarchical hybrid Ni@CoxSy/poly(3,4-ethylenedioxythiophene)-reduced graphene oxide (Ni@CoxSy/PEDOT-rGO) is rationally designed and in situ facilely synthesized as electrocatalyst to construct a binder-free sensing platform for non-enzymatic glucose monitoring through traditional electrodeposition procedure. The as-prepared Ni@CoxSy/PEDOT-rGO presents unique hierarchical structure and multiple valence states as well as strong and robust adhesion between Ni@CoxSy/PEDOT-rGO and GCE. Profiting from the aforementioned merits, the sensing platform constructed under optimal conditions achieved a wide detection range (0.2 µM ~ 2.0 mM) with high sensitivity (1546.32 µA cm-2 mM-1), a rapid response time (5 s), an ultralow detection limit (0.094 µM), superior anti-interference performance, excellent reproducibility and considerable stability. Furthermore, the sensor demonstrates an acceptable accuracy and appreciable recoveries ranging from 90.0 to 102.0% with less than 3.98% RSD in human blood serum samples, indicating the prospect of the sensor for the real samples analysis. It will provide a strategy to rationally design and fabricate ternary hierarchical hybrid as nanozyme for glucose assay.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Polímeros / Glucemia / Cobalto / Compuestos Bicíclicos Heterocíclicos con Puentes / Grafito / Níquel Límite: Humans Idioma: En Revista: Mikrochim Acta Año: 2024 Tipo del documento: Article País de afiliación: China

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Polímeros / Glucemia / Cobalto / Compuestos Bicíclicos Heterocíclicos con Puentes / Grafito / Níquel Límite: Humans Idioma: En Revista: Mikrochim Acta Año: 2024 Tipo del documento: Article País de afiliación: China