Comparison of Objective Visual Quality Following SMILE and SmartPulse Technology-Assisted TransPRK at a 1,050-Hz Ablation Frequency for Moderate-to-High Myopia.
J Refract Surg
; 40(7): e490-e498, 2024 Jul.
Article
en En
| MEDLINE
| ID: mdl-39007811
ABSTRACT
PURPOSE:
To compare the objective visual quality of moderate-to-high myopia corrected by small incision lenticule extraction (SMILE) and transepithelial photorefractive keratectomy (TransPRK) at a 1,050-Hz ablation frequency, assisted by Smart-Pulse technology (SCHWIND eye-tech-solutions).METHODS:
This study involved 123 patients (123 eyes) with moderate-to-high myopia between July 2020 and January 2021. They were categorized into the SMILE group (67 patients, 67 eyes) and the TransPRK group (56 patients, 56 eyes). Follow-ups were conducted at 6 months postoperatively to record the logarithm of the minimum angle of resolution visual acuity, and the Strehl ratio and higher order aberrations were measured using the Sirius anterior segment analysis device (SCHWIND eye-tech-solutions) under a 6-mm pupil diameter at various postoperative intervals.RESULTS:
At 1 week and 1 month postoperatively, the uncorrected distance visual acuity (UDVA) in the SMILE group was superior to that in the TransPRK group (P < .05 for both). At 1 week and 1 month postoperatively, the Strehl ratio value in the SMILE group was higher than that in the TransPRK group (P < .05 for both). At 1, 3, and 6 months postoperatively, coma was greater in the SMILE group than in the TransPRK group (P < .05 for all). Spherical aberrations were lower in the SMILE group than in the TransPRK group at 3 and 6 months postoperatively (P < .05). At 6 months postoperatively, UDVA was -0.09 ± 0.08 and -0.11 ± 0.05 logMAR in the SMILE and TransPRK groups, respectively, which exceeded their preoperative corrected distance visual acuity of -0.05 ± 0.04 and -0.09 ± 0.08 logMAR (all P < .001). Compared with preoperative values, the Strehl ratio, total higher order, coma, and spherical aberration differences were significantly increased postoperatively in both groups (all P < .001).CONCLUSIONS:
Both surgical methods improved UDVA and each had its advantages. The visual quality of SMILE was superior at 1 week and 1 month postoperatively (Strehl ratio values were higher than those of the TransPRK group), and its spherical aberration was lower than that of the TransPRK group at 3 and 6 months; TransPRK with SmartPulse technology with a 1,050-Hz ablation frequency showed that coma was significantly lower than that of the SMILE group at 1, 3, and 6 months postoperatively. [J Refract Surg. 2024;40(7)e490-e498.].
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Refracción Ocular
/
Agudeza Visual
/
Queratectomía Fotorrefractiva
/
Sustancia Propia
/
Láseres de Excímeros
Límite:
Adult
/
Female
/
Humans
/
Male
Idioma:
En
Revista:
J Refract Surg
Asunto de la revista:
OFTALMOLOGIA
Año:
2024
Tipo del documento:
Article