Smart Bioimpedance Device for the Assessment of Peripheral Muscles in Patients with COPD.
Naranjo-Hernández, David; Reina-Tosina, Javier; Roa, Laura M; Barbarov-Rostán, Gerardo; Ortega-Ruiz, Francisco; Cejudo Ramos, Pilar.
Afiliación
  • Naranjo-Hernández D; Biomedical Engineering Group, Department of Signal Theory and Communications, University of Seville, 41092 Seville, Spain.
  • Reina-Tosina J; Biomedical Engineering Group, Department of Signal Theory and Communications, University of Seville, 41092 Seville, Spain.
  • Roa LM; Biomedical Engineering Group, Department of Signal Theory and Communications, University of Seville, 41092 Seville, Spain.
  • Barbarov-Rostán G; Biomedical Engineering Group, Department of Signal Theory and Communications, University of Seville, 41092 Seville, Spain.
  • Ortega-Ruiz F; Medical Surgical Unit of Respiratory Diseases, Virgen del Rocío University Hospital, 41013 Seville, Spain.
  • Cejudo Ramos P; Medical Surgical Unit of Respiratory Diseases, Virgen del Rocío University Hospital, 41013 Seville, Spain.
Sensors (Basel) ; 24(14)2024 Jul 17.
Article en En | MEDLINE | ID: mdl-39066045
ABSTRACT
Muscle dysfunction and muscle atrophy are common complications resulting from Chronic Obstructive Pulmonary Disease (COPD). The evaluation of the peripheral muscles can be carried out through the assessment of their structural components from ultrasound images or their functional components through isometric and isotonic strength tests. This evaluation, performed mainly on the quadriceps muscle, is not only of great interest for diagnosis, prognosis and monitoring of COPD, but also for the evaluation of the benefits of therapeutic interventions. In this work, bioimpedance spectroscopy technology is proposed as a low-cost and easy-to-use alternative for the evaluation of peripheral muscles, becoming a feasible alternative to ultrasound images and strength tests for their application in routine clinical practice. For this purpose, a laboratory prototype of a bioimpedance device has been adapted to perform segmental measurements in the quadriceps region. The validation results obtained in a pseudo-randomized study in patients with COPD in a controlled clinical environment which involved 33 volunteers confirm the correlation and correspondence of the bioimpedance parameters with respect to the structural and functional parameters of the quadriceps muscle, making it possible to propose a set of prediction equations. The main contribution of this manuscript is the discovery of a linear relationship between quadriceps muscle properties and the bioimpedance Cole model parameters, reaching a correlation of 0.69 and an average error of less than 0.2 cm regarding the thickness of the quadriceps estimations from ultrasound images, and a correlation of 0.77 and an average error of 3.9 kg regarding the isometric strength of the quadriceps muscle.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Impedancia Eléctrica / Enfermedad Pulmonar Obstructiva Crónica Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Sensors (Basel) Año: 2024 Tipo del documento: Article País de afiliación: España

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Impedancia Eléctrica / Enfermedad Pulmonar Obstructiva Crónica Límite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Sensors (Basel) Año: 2024 Tipo del documento: Article País de afiliación: España