Your browser doesn't support javascript.
loading
Attogram detection limit for aqueous dye samples by laser-induced fluorescence.
Science ; 219(4586): 845-7, 1983 Feb 18.
Article en En | MEDLINE | ID: mdl-6823553
A modified flow cytometer has been used to detect attogram quantities of aqueous rhodamine 6G by laser-induced fluorescence analysis. A detection limit of 28 attograms (35,000 molecules) was obtained, nearly two orders of magnitude better than earlier measurements. The detection limit in concentration units was 1.4 x 10(-13) mole per liter. During the 1-second measurement period, the total volume sampled was 0.42 microliter. On average, only half a rhodamine 6G molecule was present in the 6-picoliter probed volume.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Rodaminas / Xantenos / Citometría de Flujo Tipo de estudio: Diagnostic_studies Idioma: En Revista: Science Año: 1983 Tipo del documento: Article
Buscar en Google