Multiple drug hemodynamic control by means of a supervisory-fuzzy rule-based adaptive control system: validation on a model.
Held, C M; Roy, R J.
Afiliación
  • Held CM; Biomedical Engineering Department, Rensselaer Polytechnic Institute, Troy, NY 12180-3590, USA.
IEEE Trans Biomed Eng ; 42(4): 371-85, 1995 Apr.
Article en En | MEDLINE | ID: mdl-7729836
ABSTRACT
A control device that uses an expert system approach for a two input-two output system has been developed and evaluated using a mathematical model of the hemodynamic response of a dog. The two inputs are the infusion rates of two drugs sodium nitroprusside (SNP) and dopamine (DPM). The two controlled variables are the mean arterial pressure and the cardiac output. The control structure is dual mode, i.e., it has two levels a critical conditions (coarse) control mode and a noncritical conditions (fine) control mode. The system switches from one to the other when threshold conditions are met. Different "controller parameters sets"-including the values for the threshold conditions-can be given to the system which will lead to different controller outputs. Both control modes are rule-based, and supervisory capabilities are added to ensure adequate drug delivery. The noncritical control mode is a fuzzy logic controller. The system includes heuristic features typically considered by anesthesiologists, like waiting periods and the observance of a "forbidden dosage range" for DPM infusion when used as an inotrope. An adaptation algorithm copes with the wide range of sensitivities to SNP found among different individuals, as well as the time varying sensitivity frequently observed in a single patient. The control device is eventually tested on a nonlinear model, designed to mimic the conditions of congestive heart failure in a dog. The test runs show a highest overshoot of 3 mmHg with nominal SNP sensitivity. When tested with different simulated SNP sensitivities, the controller adaptation produces a faster response to lower sensitivities, and reduced oscillations to higher sensitivities. The simulations seem to show that the system is able to drive and adequately keep the two hemodynamic variables within prescribed limits.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Presión Sanguínea / Nitroprusiato / Gasto Cardíaco / Dopamina / Lógica Difusa / Modelos Cardiovasculares Tipo de estudio: Diagnostic_studies / Evaluation_studies / Prognostic_studies Límite: Animals Idioma: En Revista: IEEE Trans Biomed Eng Año: 1995 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Presión Sanguínea / Nitroprusiato / Gasto Cardíaco / Dopamina / Lógica Difusa / Modelos Cardiovasculares Tipo de estudio: Diagnostic_studies / Evaluation_studies / Prognostic_studies Límite: Animals Idioma: En Revista: IEEE Trans Biomed Eng Año: 1995 Tipo del documento: Article País de afiliación: Estados Unidos