1.
Rev Sci Instrum ; 81(10): 10D916, 2010 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-21033948

RESUMO

We developed and installed an electron cyclotron emission radiometer for taking measurements of Korea Superconducting Tokamak Advanced Research (KSTAR) plasma. In order to precisely measure the absolute value of electron temperatures, a calibration measurement of the whole radiometer system was performed, which confirmed that the radiometer has an acceptably linear output signal for changes in input temperature. It was also found that the output power level predicted by a theoretical calculation agrees with that obtained by the calibration measurement. We also showed that the system displays acceptable noise-temperature performance around 0.23 eV.

2.
Rev Sci Instrum ; 81(10): 10D919, 2010 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-21033951

RESUMO

A combined system of microwave imaging reflectometry and electron cyclotron emission (ECE) imaging has been developed for the Large Helical Device. This system includes a wide-band two-dimensional horn-antenna mixer array (HMA). The HMA consists of horn antennas, waveguides, mixers, and intermediate frequency circuits. The frequency response of the HMA is between 50 and 110 GHz. The ECE signal is selected using a 95 GHz local oscillator and a 93 GHz high-pass filter.