J Perinatol ; 37(9): 1028-1031, 2017 09.
Artigo em Inglês | MEDLINE | ID: mdl-28617429

RESUMO

OBJECTIVE: To compare the performance and accuracy of the JM-103 transcutaneous bilirubinometer and Bilistick System in measuring total serum bilirubin for the early identification of neonatal hyperbilirubinemia. STUDY DESIGN: The study was performed on 126 consecutive term and near-term (⩾36 weeks' gestational age) jaundiced newborns in Cairo University Children Hospital NICU, Egypt. Total serum bilirubin was assayed concurrently by the clinical laboratory and Bilistick System and estimated using the JM-103 transcutaneous bilirubin instrument. Bland-Altman analysis was used to evaluate the agreement between determinations. RESULT: The limits of agreement of the Bilistick System (-5.8 to 3.3 mg dl-1) and JM-103 system (-5.4 to 6.0 mg dl-1) versus the clinical laboratory results were similar. CONCLUSION: The Bilistick System is an accurate alternative to transcutaneous (TcB) determination for early diagnosis and proper management of the neonatal jaundice.


Assuntos
Bilirrubina/sangue , Icterícia Neonatal/sangue , Triagem Neonatal/métodos , Biomarcadores/sangue , Egito , Feminino , Idade Gestacional , Humanos , Recém-Nascido , Icterícia Neonatal/diagnóstico , Masculino