1.
Appl Opt ; 40(36): 6677-81, 2001 Dec 20.
Artigo em Inglês | MEDLINE | ID: mdl-18364979

RESUMO

A unique scheme has been applied for sensitive remote detection of 2,4,6-trinitrotoluene (TNT) vapor trace amounts at atmospheric pressure and 24 degrees C. The detection concept is based on a single laser beam inducing a tandem process: photodissociation of TNT vapor followed by highly selective detection of its photofragments vibrationally excited NO, utilizing laser-induced fluorescence with the A2Sigma+(v' = 0) <-- X2Pi(v'' = 2) transition. A detection sensitivity of at least 8 parts in 10(9) of TNT vapor with a signal-to-noise ratio of approximately 10 has been experimentally verified for an unfocused approximately 5-mJ laser beam, measured at a distance of approximately 15 cm from the TNT sample.