Appl Radiat Isot ; 100: 60-4, 2015 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-25481523

RESUMO

In this work we compare the kVp estimate between CDTN multipurpose instrument, UnforsXI and Radcal 4075 meters under different combinations of voltage and filtration. The non-invasively measurements made using x-ray diagnostic and interventional radiology devices show similar tendencies to increase the kVp estimate when aluminum filters are placed in the path of the x-ray beam. The results reveal that the kVp estimate made by the CDTN multipurpose instrument is always satisfactory for highly filtered beam intensities.