Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Phys Med Biol ; 68(24)2023 Dec 12.
Artigo em Inglês | MEDLINE | ID: mdl-37988759

RESUMO

In recent years Radiotherapy linear accelerator (linac) vendors have developed their own integrated quality control (QC) systems. Such manufacturer-integrated-quality-control (MIQC) has the potential to improve both the quality and efficiency of linac QC but is currently being developed and utilised in the absence of specific best-practice guidance. An Institute of Physics and Engineering in Medicine working party was commissioned with a view to develop guidance for the commissioning and implementation of MIQC. This study is based upon a survey of United Kingdom (UK) radiotherapy departments performed by the working party. The survey was distributed to all heads of radiotherapy physics in the UK and investigated availability and uptake, community beliefs and opinions, utilisation, user experience and associated procedures. The survey achieved a 95% response rate and demonstrated strong support (>95%) for its use and further development. MIQC systems are available in 79% of respondents' centres, and are in clinical use in 66%. The most common MIQC system was Varian MPC, in clinical use in 58% of responding centres, with CyberKnife AQA\E2E in 11%, TomoTherapy TQA in 8% and no users of Elekta Machine QA. A majority of users found their MIQC to be easy to use, reliable, and had five or more years of experience. Most users reported occasions of discrepancy in results between MIQC and conventional testing, but the majority considered this acceptable, indicating a false reporting frequency of quarterly or less. MIQC has shown value in preventative maintenance and early detection of machine deviations. There were inconsistent approaches in the utilisation and commissioning tests performed. Fewer than half of users perform QC of MIQC. 45% of responders have modified their QC processes with the introduction of MIQC, via replacement of conventional tests or reduction in their frequency. Future guidance is recommended to assist in the implementation of MIQC.


Assuntos
Aceleradores de Partículas , Planejamento da Radioterapia Assistida por Computador , Planejamento da Radioterapia Assistida por Computador/métodos , Inquéritos e Questionários , Controle de Qualidade , Reino Unido
2.
Phys Med Biol ; 61(23): 8360-8394, 2016 12 07.
Artigo em Inglês | MEDLINE | ID: mdl-27819253

RESUMO

Flattening filter free (FFF) beams are now commonly available with new standard linear accelerators. These beams have recognised clinical advantages in certain circumstances, most notably the reduced beam-on times for high dose per fraction stereotactic treatments. Therefore FFF techniques are quickly being introduced into clinical use. The purpose of this report is to provide practical implementation advice and references for centres implementing FFF beams clinically. In particular UK-specific guidance is given for reference dosimetry and radiation protection.


Assuntos
Filtração/normas , Aceleradores de Partículas/normas , Guias de Prática Clínica como Assunto/normas , Proteção Radiológica/normas , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada/normas , Filtração/instrumentação , Humanos , Aceleradores de Partículas/instrumentação , Proteção Radiológica/instrumentação , Radiometria/métodos , Radioterapia de Intensidade Modulada/instrumentação , Reino Unido