1.
Appl Radiat Isot ; 70(2): 388-91, 2012 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-22055844

RESUMO

Two secondary standard systems of beta radiation were used to calibrate a PTW extrapolation chamber Model 23391. Three (90)Sr+(90)Y sources of different activities were used in this calibration procedure. Medium-term stability of the response of the chamber was also studied. The calibration was performed with and without field-flattening filters. The relative standard deviation of the obtained calibration factors was 8.3% for the aluminum collecting electrode and 4.1% for the graphite collecting electrode.


Assuntos
Radiometria/métodos , Radioisótopos de Estrôncio/normas , Radioisótopos de Ítrio/normas , Alumínio , Partículas beta , Calibragem , Eletrodos , Filtração/instrumentação , Grafite , Radiometria/normas , Reprodutibilidade dos Testes