1.
J Synchrotron Radiat
; 24(Pt 4): 781-786, 2017 Jul 01.
Artigo
em Inglês
| MEDLINE
| ID: mdl-28664885
RESUMO
A method to calibrate and stabilize the incident X-ray energy for anomalous diffraction data collection is provided and has been successfully used at the single-crystal diffraction beamline 1W2B at the Beijing Synchrotron Radiation Facilities. Employing a feedback loop to control the movement of the double-crystal monochromator, this new method enables the incident X-ray energy to be kept within a 0.2â eV range at the inflection point of the absorption edge.