Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
EuroIntervention ; 8(5): 556-62, 2012 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-22995081

RESUMO

AIMS: High rates of permanent pacemaker (PPM) implantation are reported after transcatheter aortic valve implantation (TAVI) using the Medtronic CoreValve® system. The Accutrak™ catheter is designed to allow a more predictable landing zone. Little is known about the real clinical impact of this catheter. The aims of this paper were to describe the potential impact of the Accutrak™ catheter on the accuracy of positioning a 26 or 29 mm CoreValve® across the aortic annulus and its impact on the need for a pacemaker. METHODS AND RESULTS: A total of 134 patients were treated with the CoreValve® Accutrak™ system at two French centres (Lille and Toulouse). Mean age was 82.4 ± 4.7 years; logistic EuroSCORE was 24.3 ± 9.5%. Procedural success rate was 99.2%; mean depth of implantation was 4.9 mm. A final position between 0 and 6 mm was achieved in 85.8% of the patients. All-cause mortality at 30 days was 6%. The PPM implantation rate was 10.6%. Due to a limited number of events, we could not identify any predictor of need for a PPM: pre-existing right bundle branch block (RBBB) (OR 2.72 [0.63-11.87], p=ns), use of a 29 mm prosthesis (OR 2.73 [0.33-22.90], p=ns) and left ventricular septal hypertrophy (OR 2.63 [0.08-83.32], p=ns). CONCLUSIONS: In this cohort of patients treated with the CoreValve® Accutrak™ system, the incidence of permanent pacemaker implantation was low, which may be a consequence of an average small implantation depth. The Accutrak™ catheter seems to be helpful in achieving higher and more predictable implants. Operators could standardise their technique to place the CoreValve® prostheses less than 6 mm below the aortic annulus.


Assuntos
Valva Aórtica/cirurgia , Catéteres , Implante de Prótese de Valva Cardíaca/métodos , Marca-Passo Artificial , Idoso , Idoso de 80 Anos ou mais , Cateterismo Cardíaco , Feminino , Implante de Prótese de Valva Cardíaca/instrumentação , Humanos , Masculino , Resultado do Tratamento
2.
J Invasive Cardiol ; 24(7): 342-4, 2012 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-22781474

RESUMO

Transcatheter aortic valve implantation (TAVI) is associated with a risk of coronary obstruction. This complication is potentially lethal when the origin of the coronary arteries is anomalous. We describe two cases of TAVI with the SAPIEN XT (Edwards Lifesciences) and CoreValve devices (Medtronic) in patients with a single coronary artery. The tools and techniques used to anticipate the risk of acute coronary occlusion are discussed.


Assuntos
Estenose da Valva Aórtica/cirurgia , Valva Aórtica/cirurgia , Cateteres Cardíacos , Oclusão Coronária/epidemiologia , Anomalias dos Vasos Coronários/complicações , Próteses Valvulares Cardíacas/classificação , Implantação de Prótese/métodos , Idoso de 80 Anos ou mais , Angiografia Coronária , Anomalias dos Vasos Coronários/diagnóstico por imagem , Feminino , Próteses Valvulares Cardíacas/efeitos adversos , Humanos , Implantação de Prótese/efeitos adversos , Implantação de Prótese/instrumentação , Fatores de Risco , Seio Aórtico/anormalidades , Resultado do Tratamento
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...