Appl Radiat Isot ; 64(10-11): 1340-5, 2006.
Artigo em Inglês | MEDLINE | ID: mdl-16600602

RESUMO

The ETNA software has been developed to compute efficiency transfer and coincidence summing corrections. Different experiments are combined to test the validity of this last facility. Point sources with multi-gamma emitters are measured at several source-to-detector distances. Experimental correction factors are determined from the variation in the peaks' relative intensities versus the geometrical conditions. The ETNA code is used to compute the corrections due to coincidence summing for the same geometries. The calculated values are compared to the experimental ones.


Assuntos
Algoritmos , Artefatos , Radioisótopos/análise , Radiometria/métodos , Validação de Programas de Computador , Software , Espectrometria gama/métodos , Simulação por Computador , Raios gama , Modelos Químicos , Modelos Moleculares , Modelos Estatísticos , Método de Monte Carlo , Doses de Radiação , Reprodutibilidade dos Testes , Sensibilidade e Especificidade