Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Clin Infect Dis ; 64(3): 335-342, 2017 Feb 01.
Artigo em Inglês | MEDLINE | ID: mdl-27927870

RESUMO

BACKGROUND: An urgent UK investigation was launched to assess risk of invasive Mycobacterium chimaera infection in cardiothoracic surgery and a possible association with cardiopulmonary bypass heater-cooler units following alerts in Switzerland and The Netherlands. METHODS: Parallel investigations were pursued: (1) identification of cardiopulmonary bypass-associated M. chimaera infection through national laboratory and hospital admissions data linkage; (2) cohort study to assess patient risk; (3) microbiological and aerobiological investigations of heater-coolers in situ and under controlled laboratory conditions; and (4) whole-genome sequencing of clinical and environmental isolates. RESULTS: Eighteen probable cases of cardiopulmonary bypass-associated M. chimaera infection were identified; all except one occurred in adults. Patients had undergone valve replacement in 11 hospitals between 2007 and 2015, a median of 19 months prior to onset (range, 3 months to 5 years). Risk to patients increased after 2010 from <0.2 to 1.65 per 10000 person-years in 2013, a 9-fold rise for infections within 2 years of surgery (rate ratio, 9.08 [95% CI, 1.81-87.76]). Endocarditis was the most common presentation (n = 11). To date, 9 patients have died. Investigations identified aerosol release through breaches in heater-cooler tanks. Mycobacterium chimaera and other pathogens were recovered from water and air samples. Phylogenetic analysis found close clustering of strains from probable cases. CONCLUSIONS: We identified low but escalating risk of severe M. chimaera infection associated with heater-coolers with cases in a quarter of cardiothoracic centers. Our investigations strengthen etiological evidence for the role of heater-coolers in transmission and raise the possibility of an ongoing, international point-source outbreak. Active management of heater-coolers and heightened clinical awareness are imperative given the consequences of infection.


Assuntos
Ponte Cardiopulmonar/efeitos adversos , Contaminação de Equipamentos , Infecções por Mycobacterium não Tuberculosas/epidemiologia , Micobactérias não Tuberculosas/isolamento & purificação , Equipamentos Cirúrgicos/microbiologia , Infecção da Ferida Cirúrgica/epidemiologia , Infecção da Ferida Cirúrgica/microbiologia , Adulto , Idoso , Idoso de 80 Anos ou mais , Microbiologia do Ar , Criança , Estudos de Coortes , Feminino , Humanos , Masculino , Pessoa de Meia-Idade , Infecções por Mycobacterium não Tuberculosas/microbiologia , Infecções por Mycobacterium não Tuberculosas/mortalidade , Infecções por Mycobacterium não Tuberculosas/transmissão , Micobactérias não Tuberculosas/classificação , Micobactérias não Tuberculosas/genética , Estudos Retrospectivos , Fatores de Risco , Infecção da Ferida Cirúrgica/mortalidade , Reino Unido/epidemiologia , Microbiologia da Água
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...