Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 7 de 7
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
J Fam Pract ; 71(4): E9-E11, 2022 05.
Artigo em Inglês | MEDLINE | ID: mdl-35730707

RESUMO

This case highlights an impressive manifestation of a diagnosis that affects many people around the world.

2.
J Cardiovasc Comput Tomogr ; 15(3): 240-245, 2021.
Artigo em Inglês | MEDLINE | ID: mdl-32868247

RESUMO

BACKGROUND: Coronary artery calcium (CAC) scanning is commonly performed before coronary CT angiography (CTA) based partly on its potential to influence CTA scan parameters. Encompassing the whole heart and performed at high tube potential (120 â€‹kVp), standard (Agatston) CAC scanning adds to patient radiation exposure. Most CAC exists in the proximal and mid coronary segments and is easily visualized at low kVp. METHODS: We tested the impact of a modified calcium scan on coronary CTA acquisition decision-making and image quality in a randomized clinical trial. Providers documented planned CTA acquisition parameters prior to CAC scanning in a blinded manner. Standard Agatston CAC scans proceeded in typical fashion whereas modified scans utilized 80 â€‹kVp and reduced z-axis length focused on the proximal-to-mid coronary arteries. CTA providers reviewed the CAC burden then documented final acquisition parameters. RESULTS: The study included 172 patients (48% female; mean age 59 â€‹± â€‹6.7). As planned, the calcium scan effective dose was significantly lower in the modified CAC scan group (0.14 vs. 0.74 â€‹mSv using a 0.014 k-factor or 0.26 vs. 1.38 â€‹mSv using a 0.026 k-factor; both p â€‹< â€‹0.001). Initially selected CTA parameters were changed at an identical rate following visual CAC assessment (59%). There was no significant difference in coronary CTA image quality (median quality score â€‹= â€‹4 in both groups, p â€‹= â€‹0.26), noise (31.0 vs 31.4 HU; p â€‹= â€‹0.81), or signal/noise ratio (17.9 vs 16.8; p â€‹= â€‹0.26). CONCLUSIONS: A low-kVp scan with focused field-of-view provides actionable information regarding the presence and severity of CAC prior to coronary CTA. Coronary CTA parameters based on patient variables are frequently modified after assessing CAC burden in the CTA suite. CLINICALTRIALS. GOV REGISTRATION NUMBER: NCT02972242.


Assuntos
Angiografia por Tomografia Computadorizada , Angiografia Coronária , Doença da Artéria Coronariana/diagnóstico por imagem , Calcificação Vascular/diagnóstico por imagem , Idoso , Feminino , Humanos , Masculino , Maryland , Pessoa de Meia-Idade , Valor Preditivo dos Testes , Estudos Prospectivos , Doses de Radiação , Exposição à Radiação , Índice de Gravidade de Doença
6.
Mil Med ; 183(5-6): e272-e275, 2018 05 01.
Artigo em Inglês | MEDLINE | ID: mdl-29415142

RESUMO

This case demonstrates the complementary benefit of utilizing multimodality cardiac imaging in the assessment of myocardial infarction with nonobstructive coronary artery disease especially when a culprit lesion is not discovered upon initial coronary catheterization. Use of cardiac magnetic resonance imaging, optical coherence tomography, and invasive coronary angiography together solidified the diagnosis of unstable, complex coronary artery disease in this case.


Assuntos
Angiografia Coronária/métodos , Doença da Artéria Coronariana/diagnóstico , Ecocardiografia/métodos , Dor no Peito/diagnóstico por imagem , Dor no Peito/etiologia , Doença da Artéria Coronariana/diagnóstico por imagem , Eletrocardiografia/métodos , Humanos , Masculino , Pessoa de Meia-Idade , Militares
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA