Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Appl Radiat Isot ; 83 Pt A: 8-11, 2014 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-24215812

RESUMO

Cancer treatments such as intensity-modulated radiotherapy (IMRT) require increasingly complex methods to verify the accuracy and precision of the treatment delivery. In vivo dosimetry based on measurements made in an electronic portal imaging device (EPID) has been demonstrated. The distorting effect of the patient anatomy on the beam intensity means it is difficult to separate changes in patient anatomy from changes in the beam intensity profile. Alternatively, upstream detectors scatter and attenuate the beam, changing the energy spectrum of the beam, and generate contaminant radiation such as electrons. We used the VANILLA device, a Monolithic Active Pixel Sensor (MAPS), to measure the 2D beam profile of a 6 MV X-ray beam at Bristol Hospital in real-time in an upstream position to the patient without clinically significant disturbance of the beam (0.1% attenuation). MAPSs can be made very thin (~20 µm) with still a very good signal-to-noise performance. The VANILLA can reconstruct the collimated beam edge with approximately 64 µm precision.


Assuntos
Radiometria/instrumentação , Radioterapia de Intensidade Modulada/instrumentação , Desenho de Equipamento , Humanos , Neoplasias/radioterapia , Fótons , Radiometria/estatística & dados numéricos , Radioterapia de Alta Energia/instrumentação , Radioterapia de Alta Energia/estatística & dados numéricos , Radioterapia de Intensidade Modulada/estatística & dados numéricos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...