Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
J Arthroplasty ; 39(5): 1335-1340, 2024 May.
Artigo em Inglês | MEDLINE | ID: mdl-37977306

RESUMO

BACKGROUND: Sequential modifications to the manufacturing process of highly cross-linked polyethylene (HXLPE) have improved the wear resistance and implant survival of these liners in total hip arthroplasty (THA). However, no study has examined the long-term (mean 10 year) wear rates and clinical outcomes of third-generation HXLPE in THA. The aim of our study was to report the longest-to-date analysis of wear rates and clinical outcomes of a third-generation HXLPE liner. METHODS: A series of 133 THAs using a specific HXLPE acetabular liner were retrospectively evaluated. Linear and volumetric wear rates were determined using a validated radiographic technique and clinical outcomes were analyzed. Multivariate analyses were performed to determine risk factors for accelerated wear. RESULTS: At a mean follow-up of 10.4 years (range, 8 to 13.4), the mean linear wear rate was 0.0172 mm/year and the mean volumetric wear rate was 16.99 mm3/year. There were no instances of osteolysis or mechanical failures at any time point and there was a 100% acetabular component survival rate. Younger age and use of offset liners were independent risk factors for increased wear (P < .01 for both). CONCLUSIONS: Our series of a third-generation HXLPE demonstrated very low wear rates and excellent implant survival at a mean of 10.4 years following primary THA. Future comparative studies at the 15- and 20-year follow-up timepoints are necessary to determine if such findings translate to true improvements in the tribological properties and longevity of these liners when compared to previous generations of HXLPE liners.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...