RESUMO
The aim of this study was the development and analytically validation of a radioimmunoassay system for the measurement of the serum ferritin concentration as one of the laboratory biomarkers for infection by Covid-19. The main components of the system were prepared in our laboratories. The first component ferritin was extracted and purified from human spleen with high purity. The second component was the 125I-labelled ferritin tracer, prepared using Chloramine-T method. Furthermore anti-ferritin antibodies and ferritin standards were provided. The developed system is sensitive, precise, reproducible and. can be translated into a kit formulation suitable for measuring serum ferritin for the detection of Covid-19 in patients at low costs and high efficiency.
RESUMO
This study aimed to provide an easy and effective method for extraction and purification of prostate-specific antigen (PSA) from human seminal fluid with high quantity (14 mg) and high purity (98%). The obtained PSA was injected into rabbits for production of anti-PSA polyclonal antibody (titer 1/1000), labeled with radioactive iodine-125 for preparation of radioactive PSA tracer (purity 98 ± 1.8% and specific activity 64 ± 1.9 µCi/µg), and used in preparation of PSA standards. All prepared components can be used in PSA immunoassays specially radioimmunoassay (RIA) kit preparation as a diagnostic tool for prostatic diseases.