1.
Invest Radiol ; 29(6): 648-51, 1994 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-8088975

RESUMO

RATIONALE AND OBJECTIVES: Easy dislodgment and a pneumothorax rate of up to 50% have been reported with the use of mammographic hookwires for radiologic guidance in thoracoscopic surgery. An alternative design is described and preliminary in vitro results are reported. METHODS: The new design is based on the T-fastener used in percutaneous gastrostomies and can be deposited with a 20-Fr guidance needle. In vitro measurements of the anchoring capability of this design were compared with the anchoring capability of the Hawkin's III (Meditech/Boston Scientific, Watertown, MA) mammographic hook wire system. RESULTS: Anchoring capabilities of the alternative anchoring design and the Hawkin's III mammographic hook wire localization system are comparable. Complete dislodgement of both anchors occurs at approximately 350 gm. CONCLUSIONS: Preliminary in vitro investigation demonstrated similar anchoring capabilities of the new design and mammographic hook wires. Potential advantages of this suture-based localization device over hook wires are discussed.


Assuntos
Pulmão/cirurgia , Radiografia Intervencionista/instrumentação , Cirurgia Torácica/instrumentação , Desenho de Equipamento , Humanos , Modelos Estruturais , Agulhas , Suturas , Toracoscopia