1.
J Bras Pneumol ; 38(2): 214-7, 2012.
Artigo em Inglês, Português | MEDLINE | ID: mdl-22576430

RESUMO

We aimed to test a new model of self-expanding tracheal stent so that it might be made available for clinical use. Using direct laryngoscopy, we placed polyurethane-coated, nitinol stents into the middle third of the trachea in 25 New Zealand rabbits. After a mean observation period of 26 days, we evaluated stent migration, degree of expansion, attachment, adherence, formation of granulation tissue, presence of inflammatory infiltrate, parietal involvement, and epithelial lining. The results showed complete radial expansion, little adherence to the tracheal mucosa, and low tissue attachment, as well as high rates of granuloma formation and stent migration. This new model proved to be biocompatible and showed a behavior similar to that of other stents on the market.


Assuntos
Ligas , Materiais Revestidos Biocompatíveis/uso terapêutico , Poliuretanos , Desenho de Prótese , Stents , Estenose Traqueal/cirurgia , Animais , Brasil , Coelhos , Stents/efeitos adversos
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...