Resultados 1 - 20 de 40
Filtrar
1.
J Cardiovasc Transl Res ; 16(4): 956-967, 2023 08.
Artigo em Inglês | MEDLINE | ID: mdl-37097591

RESUMO

Coronary flow obstruction following transcatheter aortic valve-in-valve implantation (VIV-TAVI) is associated with a high mortality risk. The aim of this work was to quantify the coronary perfusion after VIV-TAVI in a high-risk aortic root anatomy. 3D printed models of small aortic root were used to simulate the implantation of a TAVI prosthesis (Portico 23) into surgical prostheses (Trifecta 19 and 21). The aortic root models were tested in a pulsatile in vitro bench setup with a coronary perfusion simulator. The tests were performed at baseline and post-VIV-TAVI procedure in aligned and misaligned commissural configurations under simulated hemodynamic rest and exercise conditions. The experimental design provided highly controllable and repeatable flow and pressure conditions. The left and right coronary mean flow did not differ significantly at pre- and post-VIV-TAVI procedure in any tested configurations. The commissural misalignment did not induce any significant alterations to the coronary flow. High-risk aortic root anatomy did not trigger coronary ostia obstruction or coronary flow alteration after transcatheter aortic valve implantation in a surgical bioprosthesis as shown from in-vitro flow loop tests.


Assuntos
Estenose da Valva Aórtica , Bioprótese , Oclusão Coronária , Implante de Prótese de Valva Cardíaca , Próteses Valvulares Cardíacas , Substituição da Valva Aórtica Transcateter , Humanos , Substituição da Valva Aórtica Transcateter/efeitos adversos , Valva Aórtica/diagnóstico por imagem , Valva Aórtica/cirurgia , Aorta Torácica/cirurgia , Falha de Prótese , Implante de Prótese de Valva Cardíaca/efeitos adversos , Implante de Prótese de Valva Cardíaca/métodos , Perfusão , Estenose da Valva Aórtica/diagnóstico por imagem , Estenose da Valva Aórtica/cirurgia , Desenho de Prótese , Resultado do Tratamento
2.
Ann Biomed Eng ; 51(4): 715-725, 2023 Apr.
Artigo em Inglês | MEDLINE | ID: mdl-36151505

RESUMO

Emerging treatments for tricuspid valve (TV) regurgitation require realistic TV pathological models for preclinical testing. The aim of this work was to investigate structural features of fresh and defrosted porcine right-heart samples as models of mild and severe functional tricuspid regurgitation (FTR) condition in ex-vivo pulsatile flow platform. Ten fresh hearts were tested ex-vivo under steady and pulsatile flow in typical right-heart loading conditions. Hemodynamics and 3D echocardiographic imaging of TV and right ventricle (RV) were acquired. Hearts were then kept frozen for 14 days, defrosted, and tested again with the same protocol. Morphometric parameters of TV and RV were derived from 3D reconstructions based on echo data. Fresh samples showed a slightly dilated TV morphology, with coaptation gaps among the leaflets. Sample freezing induced worsening of TV insufficiency, with significant (p < 0.05) increases in annulus size (annulus area and perimeter 7.7-3.1% respectively) and dilation of RV (9.5%), which led to an increase in tenting volume (123.7%). These morphologic alterations reflected into a significant increment of regurgitation fraction (27%). Together, such results suggest that fresh porcine heart samples may be a reliable ex-vivo model of mild FTR condition, which can be enhanced through freezing/thawing treatment to model a severe pathological condition.


Assuntos
Ecocardiografia Tridimensional , Insuficiência da Valva Tricúspide , Suínos , Animais , Insuficiência da Valva Tricúspide/diagnóstico por imagem , Valva Tricúspide/diagnóstico por imagem , Ventrículos do Coração
3.
IEEE Trans Biomed Circuits Syst ; 16(6): 1325-1336, 2022 12.
Artigo em Inglês | MEDLINE | ID: mdl-36260568

RESUMO

This paper presents a custom, low-cost electronic system specifically designed for rapid and quantitative detection of the malaria parasite in a blood sample. The system exploits the paramagnetic properties of malaria-infected red blood cells (iRBCs) for their magnetophoretic capture on the surface of a silicon chip. A lattice of nickel magnetic micro-concentrators embedded in a silicon substrate concentrates the iRBCs above coplanar gold microelectrodes separated by 3 µm for their detection through an impedance measurement. The sensor is designed for a differential operation to remove the large contribution given by the blood sample. The electronic readout automatically balances the sensor before each experiment and reaches a resolution of 15 ppm in the impedance measurement at 1 MHz allowing a limit of detection of 40 parasite/µl with a capture time of 10 minutes. For better reliability of the results, four sensors are acquired during the same experiment. We demonstrate that the realized platform can also detect a single infected cell in real experimental conditions, measuring human blood infected by Plasmodium falciparum malaria specie.


Assuntos
Malária , Parasitos , Animais , Humanos , Plasmodium falciparum , Impedância Elétrica , Testes de Diagnóstico Rápido , Reprodutibilidade dos Testes , Silício , Malária/diagnóstico , Malária/parasitologia , Eritrócitos
4.
Acta Biomed ; 93(4): e2022254, 2022 08 31.
Artigo em Inglês | MEDLINE | ID: mdl-36043983

RESUMO

BACKGROUND AND AIM: The aim of this study is to validate a totally non biologic training model that combines the use of ultrasound and X ray to train Urologists and Residents in Urology in PerCutaneous NephroLithotripsy (PCNL). METHODS: The training pathway was divided into three modules: Module 1, related to the acquisition of basic UltraSound (US) skill on the kidney; Module 2, consisting of correct Nephrostomy placement; and Module 3, in which a complete PCNL was performed on the model. Trainees practiced on the model first on Module 1, than in 2 and in 3. The pathway was repeated at least three times. Afterward, they rated the performance of the model and the improvement gained using a global rating score questionnaire. RESULTS: A total of 150 Urologists took part in this study. Questionnaire outcomes on this training model showed a mean 4.21 (range 1-5) of positive outcome overall. Individual constructive validity showed statistical significance between the first and the last time that trainees practiced on the PCNL model among the three different modules. Statistical significance was also found between residents, fellows and experts scores. Trainees increased their skills during the training modules. CONCLUSION: This PCNL training model allows for the acquisition of technical knowledge and skills as US basic skill, Nephrostomy placement and entire PCNL procedure. Its structured use could allow a better and safer training pathway to increase the skill in performing a PCNL.


Assuntos
Cálculos Renais , Litotripsia , Urologia , Competência Clínica , Humanos , Urologia/educação
5.
Biomed Mater ; 17(5)2022 08 23.
Artigo em Inglês | MEDLINE | ID: mdl-35944550

RESUMO

To address the need of alternatives to autologous vessels for small-calibre vascular applications (e.g. cardiac surgery), a bio-hybrid semi-degradable material composed of silk fibroin (SF) and polyurethane (Silkothane®) was herein used to fabricate very small-calibre grafts (Øin= 1.5 mm) via electrospinning. Bio-hybrid grafts werein vitrocharacterized in terms of morphology and mechanical behaviour, and compared to similar grafts of pure SF. Similarly, two native vessels from a rodent model (abdominal aorta and vena cava) were harvested and characterized. Preliminary implants were performed on Lewis rats to confirm the suitability of Silkothane® grafts for small-calibre applications, specifically as aortic insertion and femoral shunt. The manufacturing process generated pliable grafts consisting of a randomized fibrous mesh and exhibiting similar geometrical features to rat aortas. Both Silkothane® and pure SF grafts showed radial compliances in the range from 1.37 ± 0.86 to 1.88 ± 1.01% 10-2mmHg-1, lower than that of native vessels. The Silkothane® small-calibre devices were also implanted in rats demonstrating to be adequate for vascular applications; all the treated rats survived the surgery for three months after implantation, and 16 rats out of 17 (94%) still showed blood flow inside the graft at sacrifice. The obtained results lay the basis for a deeper investigation of the interaction between the Silkothane® graft and the implant site, which may deal with further analysis on the potentialities in terms of degradability and tissue formation, on longer time-points.


Assuntos
Fibroínas , Enxerto Vascular , Animais , Prótese Vascular , Poliuretanos , Ratos , Ratos Endogâmicos Lew
6.
Eur J Cardiothorac Surg ; 61(3): 647-654, 2022 Feb 18.
Artigo em Inglês | MEDLINE | ID: mdl-34363669

RESUMO

OBJECTIVES: Commissural orientation <160° is a recognized risk factor for bicuspid aortic valve repair failure. Based on this observation, repairing this subtype of aortic valve by reorienting the 2 commissures at 180° has recently been proposed. METHODS: Nine porcine hearts with aortic annulus diameters of 25 mm were selected. A pathological model of a Sievers 1 bicuspid aortic valve was obtained by suturing the coaptation line between the left and right leaflets. Each heart underwent reimplantation procedures both in the native (120°) and the reoriented (180°) configuration. After the operation, each sample was tested on a pulse duplicator at rest (heart rate 60 beats per min) and with mild exercise (heart rate 90 beats per min) conditions. RESULTS: No statistically significant difference was noted in mean and peak transvalvular aortic gradients between the 2 configurations at rest (18.6 ± 5 vs 17.5 ± 4 for the mean aortic gradient; 42.8 ± 12.7 vs 36.3 ± 5.8 for the peak aortic gradient) but the group with the 120°-oriented commissures had significantly higher mean transaortic gradients compared to the group with the 180°-oriented commissures at initial exercise stress conditions (30.1 ± 9.1 vs 24.9 ± 3.8; p value 0.002). CONCLUSIONS: The 180° commissural reorientation of the asymmetrical bicuspid aortic valve does not improve the transvalvular aortic gradient in an acute model at rest conditions, but it could do so under stress situations. Even if it is surgically more complex and time-consuming, this approach could be a good strategy to improve long-term results, particularly in young patients.


Assuntos
Insuficiência da Valva Aórtica , Doença da Válvula Aórtica Bicúspide , Doenças das Valvas Cardíacas , Animais , Valva Aórtica/patologia , Insuficiência da Valva Aórtica/cirurgia , Humanos , Estudos Retrospectivos , Suínos , Resultado do Tratamento
7.
Adv Sci (Weinh) ; 8(14): 2004101, 2021 07.
Artigo em Inglês | MEDLINE | ID: mdl-34306971

RESUMO

Malaria remains the most important mosquito-borne infectious disease worldwide, with 229 million new cases and 409.000 deaths in 2019. The infection is caused by a protozoan parasite which attacks red blood cells by feeding on hemoglobin and transforming it into hemozoin. Despite the WHO recommendation of prompt malaria diagnosis, the quality of microscopy-based diagnosis is frequently inadequate while rapid diagnostic tests based on antigens are not quantitative and still affected by non-negligible false negative/positive results. PCR-based methods are highly performant but still not widely used in endemic areas. Here, a diagnostic tool (TMek), based on the paramagnetic properties of hemozoin nanocrystals in infected red blood cells (i-RBCs), is reported on. Exploiting the competition between gravity and magnetic forces, i-RBCs in a whole blood specimen are sorted and electrically detected in a microchip. The amplitude and time evolution of the electrical signal allow for the quantification of i-RBCs (in the range 10-105 i-RBC µL-1) and the distinction of the infection stage. A preliminary validation study on 75 patients with clinical suspect of malaria shows on-field operability, without false negative and a few false positive results. These findings indicate the potential of TMek as a quantitative, stage-selective, rapid test for malaria.


Assuntos
Dispositivos Lab-On-A-Chip , Malária/diagnóstico , Eritrócitos/parasitologia , Estudos de Avaliação como Assunto , Humanos , Reprodutibilidade dos Testes , Sensibilidade e Especificidade
8.
J Cardiovasc Transl Res ; 14(3): 513-524, 2021 06.
Artigo em Inglês | MEDLINE | ID: mdl-32959169

RESUMO

Transcatheter therapies are emerging for functional mitral regurgitation (FMR) treatment, however there is lack of pathological models for their preclinical assessment. We investigated the applicability of deer hearts for this purpose.8 whole deer hearts were housed in a pulsatile flow bench. At baseline, all mitral valves featured normal coaptation. The pathological state was induced by 60-minutes intraventricular constant pressurization. It caused mitral annulus dilation (antero-posterior diameter increase from 31.8 ± 5.6 mm to 39.5 ± 4.9 mm, p = 0.001), leaflets tethering (maximal tenting height increase from 7.3 ± 2.5 mm to 12.7 ± 3.4 mm, p < 0.001) and left ventricular diameter increase (from 67.8 ± 7.5 mm to 79.4 ± 6.5 mm, p = 0.004). These geometrical reconfigurations led to restricted mitral valve leaflets motion and leaflet coaptation loss. Preliminary feasibility assessment of two FMR treatments was performed in the developed model.Deer hearts showed ability to dilate under constant pressurization and have potential to be used for realistic preclinical research of novel FMR therapies. Graphical abstract figure legend: Deer heart mitral valve fiberscopic and echocardiographic images in peak systole at baseline and after inducing the pathological conditions representing functional mitral regurgitation. In the pathological conditions lack of coaptation between the leaflets, enlargement of the antero-posterior distance (red dashed line) and the left ventricular diameter (orange dashed line) were observed.


Assuntos
Hemodinâmica , Insuficiência da Valva Mitral/fisiopatologia , Valva Mitral/fisiopatologia , Animais , Cervos , Modelos Animais de Doenças , Ecocardiografia Tridimensional , Tecnologia de Fibra Óptica , Preparação de Coração Isolado , Valva Mitral/diagnóstico por imagem , Insuficiência da Valva Mitral/diagnóstico por imagem , Função Ventricular Esquerda
9.
Sensors (Basel) ; 20(17)2020 Sep 02.
Artigo em Inglês | MEDLINE | ID: mdl-32887406

RESUMO

The development of innovative diagnostic tests is fundamental in the route towards malaria eradication. Here, we discuss the sorting capabilities of an innovative test for malaria which allows the quantitative and rapid detection of all malaria species. The physical concept of the test exploits the paramagnetic property of infected erythrocytes and hemozoin crystals, the magnetic fingerprints of malaria common to all species, which allows them to undergo a selective magnetophoretic separation driven by a magnetic field gradient in competition with gravity. Upon separation, corpuscles concentrate at the surface of a silicon microchip where interdigitated electrodes are placed in close proximity to magnetic concentrators. The impedance variation proportional to the amount of attracted particles is then measured. The capability of our test to perform the selective detection of infected erythrocytes and hemozoin crystals has been tested by means of capture experiments on treated bovine red blood cells, mimicking the behavior of malaria-infected ones, and suspensions of synthetic hemozoin crystals. Different configuration angles of the chip with respect to gravity force and different thicknesses of the microfluidic chamber containing the blood sample have been investigated experimentally and by multiphysics simulations. In the paper, we describe the optimum conditions leading to maximum sensitivity and specificity of the test.


Assuntos
Hemeproteínas , Magnetismo , Malária , Animais , Bovinos , Eritrócitos , Fenômenos Magnéticos , Malária/diagnóstico , Fenômenos Físicos
10.
Mol Biol Rep ; 46(4): 4483-4500, 2019 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-31183678

RESUMO

One of the main aims of bone tissue engineering, regenerative medicine and cell therapy is development of an optimal artificial environment (scaffold) that can trigger a favorable response within the host tissue, it is well colonized by resident cells of organism and ideally, it can be in vitro pre-colonized by cells of interest to intensify the process of tissue regeneration. The aim of this study was to develop an effective tool for regenerative medicine, which combines the optimal bone-like scaffold and colonization technique suitable for cell application. Accordingly, this study includes material (physical, chemical and structural) and in vitro biological evaluation of scaffolds prior to in vivo study. Thus, porosity, permeability or elasticity of two types of bone-like scaffolds differing in the ratio of collagen type I and natural calcium phosphate nanoparticles (bCaP) were determined, then analyzes of scaffold interaction with mesenchymal stem cells (MSCs) were performed. Simultaneously, dynamic seeding using a perfusion bioreactor followed by static cultivation was compared with standard static cultivation for the whole period of cultivation. In summary, cell colonization ability was estimated by determination of cell distribution within the scaffold (number, depth and homogeneity), matrix metalloproteinase activity and gene expression analysis of signaling molecules and differentiation markers. Results showed, the used dynamic colonization technique together with the newly-developed collagen-based scaffold with high content of bCaP to be an effective combined tool for producing bone grafts for bone implantology and regenerative medicine.


Assuntos
Fosfatos de Cálcio/metabolismo , Células-Tronco Mesenquimais/metabolismo , Engenharia Tecidual/métodos , Animais , Osso e Ossos/química , Diferenciação Celular , Células Cultivadas , Colágeno/química , Feminino , Transplante de Células-Tronco Mesenquimais/métodos , Nanopartículas , Osteogênese/efeitos dos fármacos , Medicina Regenerativa , Suínos , Alicerces Teciduais/química
11.
J Biomech Eng ; 141(10)2019 Oct 01.
Artigo em Inglês | MEDLINE | ID: mdl-31053843

RESUMO

Prosthetic valve thrombosis (PVT) is a serious complication affecting prosthetic heart valves. The transvalvular mean pressure gradient (MPG) derived by Doppler echocardiography is a crucial index to diagnose PVT but may result in false negatives mainly in case of bileaflet mechanical valves (BMVs) in mitral position. This may happen because MPG estimation relies on simplifying assumptions on the transvalvular fluid dynamics or because Doppler examination is manual and operator dependent. A deeper understanding of these issues may allow for improving PVT diagnosis and management. To this aim, we used in vitro and fluid-structure interaction (FSI) modeling to simulate the function of a real mitral BMV in different configurations: normally functioning and stenotic with symmetric and completely asymmetric leaflet opening, respectively. In each condition, the MPG was measured in vitro, computed directly from FSI simulations and derived from the corresponding velocity field through a Doppler-like postprocessing approach. Following verification versus in vitro data, MPG computational data were analyzed to test their dependency on the severity of fluid-dynamic derangements and on the measurement site. Computed MPG clearly discriminated between normally functioning and stenotic configurations. They did not depend markedly on the site of measurement, yet differences below 3 mmHg were found between MPG values at the central and lateral orifices of the BMV. This evidence suggests a mild uncertainty of the Doppler-based evaluation of the MPG due to probe positioning, which yet may lead to false negatives when analyzing subjects with almost normal MPG.

12.
Tissue Eng Part A ; 24(19-20): 1492-1503, 2018 10.
Artigo em Inglês | MEDLINE | ID: mdl-29724157

RESUMO

Tissue-engineered human blood vessels may enable in vitro disease modeling and drug screening to accelerate advances in vascular medicine. Existing methods for tissue-engineered blood vessel (TEBV) fabrication create homogenous tubes not conducive to modeling the focal pathologies characteristic of certain vascular diseases. We developed a system for generating self-assembled human smooth muscle cell (SMC) ring units, which were fused together into TEBVs. The goal of this study was to assess the feasibility of modular assembly and fusion of ring building units to fabricate spatially controlled, heterogeneous tissue tubes. We first aimed to enhance fusion and reduce total culture time, and determined that reducing ring preculture duration improved tube fusion. Next, we incorporated electrospun polymer ring units onto tube ends as reinforced extensions, which allowed us to cannulate tubes after only 7 days of fusion, and culture tubes with luminal flow in a custom bioreactor. To create focal heterogeneities, we incorporated gelatin microspheres into select ring units during self-assembly, and fused these rings between ring units without microspheres. Cells within rings maintained their spatial position along tissue tubes after fusion. Because tubes fabricated from primary SMCs did not express contractile proteins, we also fabricated tubes from human mesenchymal stem cells, which expressed smooth muscle alpha actin and SM22-α. This work describes a platform approach for creating modular TEBVs with spatially defined structural heterogeneities, which may ultimately be applied to mimic focal diseases such as intimal hyperplasia or aneurysm.


Assuntos
Vasos Sanguíneos/crescimento & desenvolvimento , Engenharia Tecidual/métodos , Aorta/citologia , Reatores Biológicos , Fusão Celular , Proliferação de Células , Células Cultivadas , Gelatina , Humanos , Cinética , Células-Tronco Mesenquimais/citologia , Microesferas , Miócitos de Músculo Liso/citologia , Poliésteres/química
13.
J Mater Sci Mater Med ; 29(3): 20, 2018 Feb 01.
Artigo em Inglês | MEDLINE | ID: mdl-29392427

RESUMO

Collagen composite scaffolds have been used for a number of studies in tissue engineering. The hydration of such highly porous and hydrophilic structures may influence mechanical behaviour and porosity due to swelling. The differences in physical properties following hydration would represent a significant limiting factor for the seeding, growth and differentiation of cells in vitro and the overall applicability of such hydrophilic materials in vivo. Scaffolds based on collagen matrix, poly(DL-lactide) nanofibers, calcium phosphate particles and sodium hyaluronate with 8 different material compositions were characterised in the dry and hydrated states using X-ray microcomputed tomography, compression tests, hydraulic permeability measurement, degradation tests and infrared spectrometry. Hydration, simulating the conditions of cell seeding and cultivation up to 48 h and 576 h, was found to exert a minor effect on the morphological parameters and permeability. Conversely, hydration had a major statistically significant effect on the mechanical behaviour of all the tested scaffolds. The elastic modulus and compressive strength of all the scaffolds decreased by ~95%. The quantitative results provided confirm the importance of analysing scaffolds in the hydrated rather than the dry state since the former more precisely simulates the real environment for which such materials are designed.


Assuntos
Colágeno/química , Dessecação , Alicerces Teciduais/química , Água/química , Materiais Biocompatíveis/química , Fosfatos de Cálcio/química , Força Compressiva , Módulo de Elasticidade , Ácido Hialurônico/química , Teste de Materiais , Fenômenos Mecânicos , Poliésteres/química , Porosidade , Engenharia Tecidual/métodos , Microtomografia por Raio-X
14.
J Biomech ; 68: 14-23, 2018 02 08.
Artigo em Inglês | MEDLINE | ID: mdl-29279196

RESUMO

The performance of blood-processing devices largely depends on the associated fluid dynamics, which hence represents a key aspect in their design and optimization. To this aim, two approaches are currently adopted: computational fluid-dynamics, which yields highly resolved three-dimensional data but relies on simplifying assumptions, and in vitro experiments, which typically involve the direct video-acquisition of the flow field and provide 2D data only. We propose a novel method that exploits space- and time-resolved magnetic resonance imaging (4D-flow) to quantify the complex 3D flow field in blood-processing devices and to overcome these limitations. We tested our method on a real device that integrates an oxygenator and a heat exchanger. A dedicated mock loop was implemented, and novel 4D-flow sequences with sub-millimetric spatial resolution and region-dependent velocity encodings were defined. Automated in house software was developed to quantify the complex 3D flow field within the different regions of the device: region-dependent flow rates, pressure drops, paths of the working fluid and wall shear stresses were computed. Our analysis highlighted the effects of fine geometrical features of the device on the local fluid-dynamics, which would be unlikely observed by current in vitro approaches. Also, the effects of non-idealities on the flow field distribution were captured, thanks to the absence of the simplifying assumptions that typically characterize numerical models. To the best of our knowledge, our approach is the first of its kind and could be extended to the analysis of a broad range of clinically relevant devices.


Assuntos
Temperatura Alta , Hidrodinâmica , Imageamento Tridimensional/instrumentação , Imageamento por Ressonância Magnética/instrumentação , Oxigenadores , Velocidade do Fluxo Sanguíneo , Humanos , Imageamento Tridimensional/métodos , Projetos Piloto , Fatores de Tempo
15.
Elife ; 62017 03 18.
Artigo em Inglês | MEDLINE | ID: mdl-28315522

RESUMO

Upon cardiac pathological conditions such as ischemia, microenvironmental changes instruct a series of cellular responses that trigger cardiac fibroblasts-mediated tissue adaptation and inflammation. A comprehensive model of how early environmental changes may induce cardiac fibroblasts (CF) pathological responses is far from being elucidated, partly due to the lack of approaches involving complex and simultaneous environmental stimulation. Here, we provide a first analysis of human primary CF behavior by means of a multi-stimulus microdevice for combined application of cyclic mechanical strain and controlled oxygen tension. Our findings elucidate differential human CFs responses to different combinations of the above stimuli. Individual stimuli cause proliferative effects (PHH3+ mitotic cells, YAP translocation, PDGF secretion) or increase collagen presence. Interestingly, only the combination of hypoxia and a simulated loss of contractility (2% strain) is able to additionally induce increased CF release of inflammatory and pro-fibrotic cytokines and matrix metalloproteinases.


Assuntos
Fibroblastos/fisiologia , Hipóxia/metabolismo , Oxigênio/metabolismo , Estresse Mecânico , Adaptação Fisiológica , Proliferação de Células , Células Cultivadas , Colágeno/metabolismo , Citocinas/metabolismo , Técnicas Citológicas/instrumentação , Humanos , Estresse Fisiológico
16.
Ann Biomed Eng ; 45(6): 1496-1510, 2017 06.
Artigo em Inglês | MEDLINE | ID: mdl-28224370

RESUMO

In the past decades, vascular tissue engineering has made great strides towards bringing engineered vascular tissues to the clinics and, in parallel, obtaining in-lab tools for basic research. Herein, we propose the design of a novel dual-mode bioreactor, useful for the fabrication (construct mode) and in vitro stimulation (culture mode) of collagen-based tubular constructs. Collagen-based gels laden with smooth muscle cells (SMCs) were molded directly within the bioreactor culture chamber. Based on a systematic characterization of the bioreactor culture mode, constructs were subjected to 10% cyclic strain at 0.5 Hz for 5 days. The effects of cyclic stimulation on matrix re-arrangement and biomechanical/viscoelastic properties were examined and compared vs. statically cultured constructs. A thorough comparison of cell response in terms of cell localization and expression of contractile phenotypic markers was carried out as well. We found that cyclic stimulation promoted cell-driven collagen matrix bi-axial compaction, enhancing the mechanical strength of strained samples with respect to static controls. Moreover, cyclic strain positively affected SMC behavior: cells maintained their contractile phenotype and spread uniformly throughout the whole wall thickness. Conversely, static culture induced a noticeable polarization of cell distribution to the outer rim of the constructs and a sharp reduction in total cell density. Overall, coupling the use of a novel dual-mode bioreactor with engineered collagen-gel-based tubular constructs demonstrated to be an interesting technology to investigate the modulation of cell and tissue behavior under controlled mechanically conditioned in vitro maturation.


Assuntos
Reatores Biológicos , Engenharia Tecidual , Células Cultivadas , Colágeno , Elasticidade , Géis , Humanos , Miócitos de Músculo Liso/fisiologia , Artéria Pulmonar/citologia , Viscosidade
17.
Ann Biomed Eng ; 45(4): 884-897, 2017 04.
Artigo em Inglês | MEDLINE | ID: mdl-27752921

RESUMO

After coronary artery bypass grafting, structural modifications of the saphenous vein wall lead to lumen narrowing in response to the altered hemodynamic conditions. Here we present the design of a novel ex vivo culture system conceived for mimicking central coronary artery hemodynamics, and we report the results of biomechanical stimulation experiments using human saphenous vein samples. The novel pulsatile system used an aortic-like pressure for forcing a time-dependent coronary-like resistance to obtain the corresponding coronary-like flow rate. The obtained pulsatile pressures and flow rates (diastolic/systolic: 80/120 mmHg and 200/100 mL/min, respectively) showed a reliable mimicking of the complex coronary hemodynamic environment. Saphenous vein segments from patients undergoing coronary artery bypass grafting (n = 12) were subjected to stimulation in our bioreactor with coronary pulsatile pressure/flow patterns or with venous-like perfusion. After 7-day stimulation, SVs were fixed and stained for morphometric evaluation and immunofluorescence. Results were compared with untreated segments of the same veins. Morphometric and immunofluorescence analysis revealed that 7 days of pulsatile stimulation: (i) did not affect integrity of the vessel wall and lumen perimeter, (ii) significantly decreased both intima and media thickness, (iii) led to partial endothelial denudation, and (iv) induced apoptosis in the vessel wall. These data are consistent with the early vessel remodeling events involved in venous bypass adaptation to arterial flow/pressure patterns. The pulsatile system proved to be a suitable device to identify ex vivo mechanical cues leading to graft adaptation.


Assuntos
Ponte de Artéria Coronária , Circulação Coronária , Modelos Cardiovasculares , Fluxo Pulsátil , Veia Safena/fisiopatologia , Humanos
18.
Ann Thorac Surg ; 103(1): 139-144, 2017 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-27544288

RESUMO

BACKGROUND: Aortic valve replacement has evolved recently with the development of the sutureless bioprosthesis. One such valve is the Perceval bioprosthesis, which is built by mounting leaflets of bovine pericardium to a thin stent; this approach has the potential to provide an excellent fluid dynamic performance. We undertook an in vitro study to compare the hydrodynamic performance of the sutureless bioprosthesis with two standard pericardial stented bioprostheses (Crown and Magna). METHODS: Tests were conducted using a mock loop, testing on two sizes of the three prostheses. The prosthesis sizes were chosen to house the valves in porcine aortic roots with a native annulus diameter of 19 mm (n = 6) or 21 mm (n = 6). The stroke volume ranged from 25 mL to 105 mL at a simulated heart rate of 70 beats per minute. RESULTS: Mean pressure drop and energy loss rose with increasing stroke volume in all of the valves tested (p < 0.001), with the sutureless valve showing the lowest values for both variables (p < 0.001). Effective orifice area values were stable across the stroke volume intervals and were larger in the sutureless valves (p < 0.001). CONCLUSIONS: All of the valves tested provided good fluid dynamic performances. The sutureless bioprosthesis provided the best performance with the least hindrance to flow behavior. From the hydrodynamic perspective, the sutureless prosthesis may present an advance in the evolution of bioprostheses, ensuring low gradients and potential for low incidence of patient-prosthesis mismatch even in small annuli.


Assuntos
Estenose da Valva Aórtica/cirurgia , Valva Aórtica/cirurgia , Bioprótese , Próteses Valvulares Cardíacas , Teste de Materiais/métodos , Pericárdio/transplante , Stents , Animais , Valva Aórtica/patologia , Estenose da Valva Aórtica/fisiopatologia , Bovinos , Hemodinâmica , Desenho de Prótese , Suínos
19.
Physiol Meas ; 37(12): 2274-2285, 2016 12.
Artigo em Inglês | MEDLINE | ID: mdl-27883333

RESUMO

Several novel approaches were recently developed to treat aortic root pathologies. The alteration induced by some of these approaches to the biomechanics of the aortic root could possibly affect the coronary perfusion, compromising the procedural outcome. In this scenario, the need to replicate in vitro the coronary flow pattern in physiological and pathological conditions is becoming crucial for the functional assessment of novel devices and techniques. This article describes the design of an easy-to-use, left-and-right coronary impedance simulator, coupled with native aortic roots for in vitro pulsatile tests. Experiments were performed in order to assess the performances of the coronary impedance simulator when coupled with healthy aortic valves (cardiac output: 3.8 ± 0.26 l min-1; mean systemic pressure: 95 ± 1.3 mmHg; mean coronary flow rate: 272 ± 13.4 ml min-1) or with regurgitant valves (cardiac output: 1.9 ± 0.24 l min-1; mean systemic pressure of 45 ± 3.3 mmHg; mean coronary flow rate:149 ± 21.9 ml min-1). The acute systemic response to valve regurgitation was also replicated, with increased beat rate and afterload, aimed at restoring the systemic pressure (cardiac output: 2.5 ± 0.23 l min-1; mean systemic pressure of 109 ± 6.1 mmHg; mean coronary flow rate: 262 ± 35.5 ml min-1). In the test conditions, the system was able to replicate in vitro the main determinants of the coronary circulation with physiological left/right coronary flow rate repartition, and a realistic interaction between coronary and systemic hemodynamics. The coronary simulator appears to be a suitable platform to study and optimize the interactions between novel approaches to aortic valve pathology and the coronary perfusion.


Assuntos
Biomimética/instrumentação , Circulação Coronária , Hemodinâmica , Aorta/fisiologia , Pressão Sanguínea , Impedância Elétrica , Desenho de Equipamento , Frequência Cardíaca , Vasoconstrição
20.
J Artif Organs ; 19(4): 350-356, 2016 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-27230086

RESUMO

Pericardial and porcine stented aortic valves have different leaflet kinematics. To study the biomechanics of a prosthesis thoroughly, the in vitro setting is the most appropriate. The aim of our study was to find out whether the prosthesis design in which the pericardial sheet is outside the stent post might influence the opening and closing patterns of the leaflets. Four pericardial prostheses (Magna Ease [MG] 21, Trifecta [TRI] 21, Soprano-Armonia [SA] 20 and Mitroflow [MF] 23) that fitted aortic roots with a native annulus diameter of 2.1 cm were implanted and their leaflet kinematics was studied by a high-speed digital camera. In the opening phase, MG showed the shortest RVOT and the highest RVOVI, with values of 12 ± 2 and 209 ± 17 ms, respectively. The RVOT of MG was significantly shorter than that of MF (p < 0.01), but not than that of TRI (p = 0.286). Both TRI and SA showed similar opening patterns (TRI: RVOT of 15 ± 3 ms and RVOVI of 132 ± 25 ms; SA: 17 ± 2 ms and 126 ± 19 ms), without statistically significant difference. Conversely, MF showed the slowest profile, with an RVOT of 23 ± 3 ms and an RVOVI of 94 ± 8 ms (Table 1; Fig. 3). The opening/closing profile is not influenced by the position of the pericardial leaflets, but depends on other intrinsic structural characteristics related to the material used for the stent and leaflets. Moreover, the kinematics does not affect the valve performance. Table 1 Kinematics and hydrodynamic results, reported as means and standard deviations, evaluated over the tested heart samples TRI SA MG MF ANOVA TRI versus SA TRI versus MG TRI versus MF SA versus MG SA versus MF MG versus MF p Value p Value p Value p Value p Value p Value p Value ET (ms) 1.0 1.0 1.0 1.0 RVOT (ms) 15 ± 3 17 ± 2 12 ± 2 23 ± 3 <0.01 1.0 0.286 <0.01 0.03 <0.01 <0.01 SVCT (ms) 247 ± 14 231 ± 15 256 ± 26 241 ± 11 0.170 0.463 0.853 0.931 0.213 1.0 1.0 RVCT (ms) 35 ± 19 52 ± 13 32 ± 17 52 ± 4 0.07 0.474 1.0 0.494 0.236 1.0 0.247 TVCT (ms) 283 ± 10 283 ± 19 289 ± 10 293 ± 11 0.584 1.00 1.0 1.0 1.0 1.0 1.0 RVOVI (ms-1) 132 ± 25 126 ± 19 209 ± 17 94 ± 8 <0.01 0.959 <0.01 0.02 <0.01 0.07 <0.01 SVCVI (ms-1) -0.9 ± 0.3 -1.1 ± 0.4 -0.57 ± 0.1 -0.55 ± 0.1 <0.01 1.0 0.353 0.292 0.045 0.04 1.0 RVCVI (ms-1) -16 ± 4 -10 ± 2 -18 ± 6 -10 ± 1 <0.01 0.396 1.0 0.513 0.025 1.0 0.03 Δp (mmHg) 6.7 ± 3.6 10.6 ± 5.5 15.2 ± 7.9 10.7 ± 6.1 <0.01 0.01 <0.01 0.01 0.04 1.0 <0.01 EOA (cm2) 2.2 ± 1.2 1.7 ± 0.9 1.5 ± 0.8 1.7 ± 0.9 <0.01 0.03 <0.01 0.01 0.261 0.617 0.11 El  % 7.3 ± 1 11.9 ± 1 15.4 ± 2 11.8 ± 3 <0.01 <0.01 <0.01 <0.01 0.04 1.00 0.03 CO (L/min) 3.1 ± 0.4 2.8 ± 0.5 3.1 ± 0.3 3.0 ± 0.5 0.534 0.282 0.792 0.702 0.106 0.552 0.559 ET ejection time, RVOT rapid valve-opening time, SVCT slow valve-closing time, RVCT rapid valve-closing time, TVCT total valve-closing time, RVOVI rapid valve-opening velocity index, SVCVI slow valve-closing velocity index, RVCVI rapid valve-closing velocity index, Δp mean pressure drop, EOA effective orifice area, El % energy loss, CO cardiac output.


Assuntos
Valva Aórtica , Bioprótese/estatística & dados numéricos , Próteses Valvulares Cardíacas/estatística & dados numéricos , Animais , Aorta , Fenômenos Biomecânicos , Débito Cardíaco , Hemodinâmica , Humanos , Hidrodinâmica , Técnicas In Vitro , Pericárdio/transplante , Desenho de Prótese , Stents , Suínos