Int J Lab Hematol ; 31(5): 496-504, 2009 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-18422712

RESUMO

The objective of the study was to determine whether selected hematologic parameters measured on umbilical cord blood samples using an automated hematology analyzer (Sysmex XE-2100) were affected by (i) anticoagulant (the specimens were collected in EDTA vs. sodium heparin), (ii) temperature (the specimens were maintained at 4 degrees C vs. room temperature for up to 72 h) and (iii) 1 : 5 dilution vs. undiluted using the manufacturer's diluting solution. Use of heparin, instead of EDTA, had little effect on the hematologic results (n = 8) except for lower platelet and progenitor cell counts. Results were remarkably stable for 72 h at either room temperature or 4 degrees C except for modest red blood cell swelling at 24 h. Specimens of blood diluted at 1 : 5 had an immediate small, but significant change on white cell count (+13.3%), reticulocyte count (-11.2%) and reticulocyte hemoglobin content (-19.6%). Diluted samples did not change further over 4 h at room temperature. With a 1 : 5 dilution, analysis of 40 microl of cord blood stored for 3 days at room temperature may provide useful hematologic information with little phlebotomy loss.


Assuntos
Anticoagulantes/farmacologia , Sangue Fetal , Testes Hematológicos/métodos , Coleta de Amostras Sanguíneas/métodos , Ácido Edético/farmacologia , Sangue Fetal/química , Sangue Fetal/efeitos dos fármacos , Heparina/farmacologia , Humanos , Recém-Nascido , Manejo de Espécimes/métodos , Temperatura
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...