Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 4 de 4
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
JACC Case Rep ; 2(5): 796-801, 2020 May.
Artigo em Inglês | MEDLINE | ID: mdl-34317350

RESUMO

We present the case of a 57-year-old man with a primary prevention internal cardioverter-defibrillator for severe nonischemic cardiomyopathy. At the time of elective replacement indicator, systolic function had fully recovered, and his generator was not changed. Nearly 5 years post-elective replacement indicator he received appropriate internal cardioverter-defibrillator therapies during a myocardial infarction. (Level of Difficulty: Intermediate.).

3.
Pacing Clin Electrophysiol ; 32(1): 134-6, 2009 Jan.
Artigo em Inglês | MEDLINE | ID: mdl-19140924

RESUMO

T-wave oversensing represents a common cause of inappropriate shocks in patients with implanted cardiac defibrillators. This case report demonstrates a strategy of device programming using V-V pace delay (sequential rather than simultaneous biventricular pacing) to eliminate T-wave oversensing without decreasing sensitivity to detect true tachyarrhythmia.


Assuntos
Desfibriladores Implantáveis/efeitos adversos , Traumatismos por Eletricidade/etiologia , Traumatismos por Eletricidade/prevenção & controle , Eletrocardiografia/efeitos adversos , Falha de Equipamento , Idoso , Feminino , Ventrículos do Coração/cirurgia , Humanos
4.
Am Heart J ; 146(2): 339-44, 2003 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-12891205

RESUMO

BACKGROUND: Removal of infected endovascular leads if often required for cure of systemic infection, but the perceived risk of embolic events in the presence of large (>10 mm) vegetations has been considered a relative contraindication to transvenous removal. Surgical removal of pacemaker leads has been suggested in this situation to avoid occurrence of pulmonary embolization. METHODS: Of 38 patients with infection of implanted pacemaker or cardioverter-defibrillator devices, those with evidence for systemic infection underwent transesophageal echocardiography to assess for the presence of vegetations. RESULTS: Vegetations on endocardial leads or right-sided cardiac structures ranging in size from 10 mm to 38 mm in their largest dimension were detected in 9 patients. All patients underwent successful transvenous removal of endocardial leads. Five of 9 patients (55%) had evidence of pulmonary embolism. However, all 5 patients made a full recovery with antibiotic treatment and anticoagulation. Among patients with endocardial vegetations, there was no difference in hospitalization periods between those with or without pulmonary embolism (14.6 +/- 0.8 days vs 18.0 +/- 4.5 days, P =.7). CONCLUSIONS: Transvenous removal of infected pacemaker leads is an alternative to open-thoracotomy removal of infected leads. Fifty-five percent of patients with vegetations on endocardial leads in our series experienced pulmonary embolism, but neither survival nor length of hospital stay were affected by this complication.


Assuntos
Desfibriladores Implantáveis/efeitos adversos , Remoção de Dispositivo/métodos , Endocardite/cirurgia , Marca-Passo Artificial/efeitos adversos , Infecções Relacionadas à Prótese/cirurgia , Adulto , Idoso , Idoso de 80 Anos ou mais , Ecocardiografia Transesofagiana , Eletrodos Implantados/efeitos adversos , Endocardite/diagnóstico por imagem , Endocardite/etiologia , Endocardite Bacteriana/diagnóstico por imagem , Endocardite Bacteriana/etiologia , Feminino , Humanos , Tempo de Internação , Masculino , Pessoa de Meia-Idade , Infecções Relacionadas à Prótese/diagnóstico por imagem , Infecções Relacionadas à Prótese/etiologia , Embolia Pulmonar/diagnóstico por imagem , Embolia Pulmonar/etiologia , Estudos Retrospectivos , Infecções Estafilocócicas/diagnóstico por imagem , Infecções Estafilocócicas/etiologia , Infecções Estafilocócicas/cirurgia , Resultado do Tratamento
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...