1.
Rev Sci Instrum ; 86(5): 055109, 2015 May.
Artigo em Inglês | MEDLINE | ID: mdl-26026560

RESUMO

Several experiments in fundamental physics demand an environment of very low, homogeneous, and stable magnetic fields. For the magnetic characterization of such environments, we present a portable SQUID system that measures the absolute magnetic flux density vector and the gradient tensor. This vector-tensor system contains 13 integrated low-critical temperature (LTc) superconducting quantum interference devices (SQUIDs) inside a small cylindrical liquid helium Dewar with a height of 31 cm and 37 cm in diameter. The achievable resolution depends on the flux density of the field under investigation and its temporal drift. Inside a seven-layer mu-metal shield, an accuracy better than ±23 pT for the components of the static magnetic field vector and ±2 pT/cm for each of the nine components of the gradient tensor is reached by using the shifting method.


Assuntos
Ambiente Controlado , Campos Magnéticos , Desenho de Equipamento , Hélio , Impressão Tridimensional , Temperatura
2.
Biomed Tech (Berl) ; 59 Suppl 1: s649-99, 2014 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-25385894
SELEÇÃO DE REFERÊNCIAS