1.
Physiol Meas ; 29(10): N59-70, 2008 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-18812644

RESUMO

Radiofrequency ablation has become an accepted treatment option of patients with primary and metastatic liver tumors. We propose an ablation electrode array consisting of 4-8 blade shaped electrodes arranged in a circular geometry for the treatment of large liver tumors. We developed a 3D code based on the finite difference method for evaluating the effect of different numbers of electrodes (4, 6 and 8) and electrode distance on lesion size. The configuration with six electrodes can ablate a volume of 70 x 70 x 40 mm(3) in approximately 5 min, with tissue temperature above 50 degrees C throughout the treatment volume. We then performed an experimental study in polyacrylamide gel in order to validate the theoretical results. The average temperature error between the simulation and the experiment was 3.8% at the center of the electrode array. This study shows that the proposed device potentially allows more rapid treatment of large tumors than current radiofrequency ablation devices.


Assuntos
Ablação por Cateter/instrumentação , Neoplasias Hepáticas/cirurgia , Animais , Bovinos , Simulação por Computador , Condutividade Elétrica , Eletrodos , Análise Numérica Assistida por Computador , Temperatura , Condutividade Térmica , Fatores de Tempo