Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
J Forensic Sci ; 65(3): 953-959, 2020 May.
Artigo em Inglês | MEDLINE | ID: mdl-31985834

RESUMO

Three commercially available integrated rapid DNA instruments were tested as a part of a rapid DNA maturity assessment in July of 2018. The assessment was conducted with sets of blinded single-source reference samples provided to participants for testing on the individual rapid platforms within their laboratories. The data were returned to the National Institute of Standards and Technology (NIST) for review and analysis. Both FBI-defined automated review (Rapid DNA Analysis) and manual review (Modified Rapid DNA Analysis) of the datasets were conducted to assess the success of genotyping the 20 Combined DNA Index System (CODIS) core STR loci and full profiles generated by the instruments. Genotype results from the multiple platforms, participating laboratories, and STR typing chemistries were combined into a single analysis. The Rapid DNA Analysis resulted in a success rate of 80% for full profiles (85% for the 20 CODIS core loci) with automated analysis. Modified Rapid DNA Analysis resulted in a success rate of 90% for both the CODIS 20 core loci and full profiles (all attempted loci per chemistry). An analysis of the peak height ratios demonstrated that 95% of all heterozygous alleles were above 59% heterozygote balance. For base-pair sizing precision, the precision was below the standard 0.5 bp deviation for both the ANDE 6C System and the RapidHIT 200.


Assuntos
Impressões Digitais de DNA/instrumentação , Impressões Digitais de DNA/normas , DNA/isolamento & purificação , Bases de Dados de Ácidos Nucleicos , Genótipo , Heterozigoto , Humanos , Repetições de Microssatélites , Mucosa Bucal/química , Controle de Qualidade
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...