Rev Sci Instrum ; 93(6): 063105, 2022 Jun 01.
Artigo em Inglês | MEDLINE | ID: mdl-35778029

RESUMO

A hyperspectral imaging system (HIS) is a helpful tool that acquires spatial and spectral information from a target. This study developed a coaxial heterogeneous HIS (CHHIS) to collect spectral images with wavelengths ranging from 400 to 1700 nm. In this system, a visible (VIS) spectrometer and a short-wave infrared (SWIR) spectrometer are combined with a coaxial optical path to share the same field of view. This structure reduces the complexity of spatial registration and maintains the scanning duration of two spectrometers as that of a single spectrometer. The spectrometers are also replaceable for extending the detecting spectral range of the system. The calibration methodologies, including spatial correction, spectral calibration, and reflectance calibration, were developed for this system. The signal-to-noise ratio of VIS and SWIR spectrometers in the CHHIS was up to 40 and 60 dB when the exposure time of the VIS and SWIR imaging sensors was 1000 and 10 ms, respectively. When the target distance was at 600 mm, the spatial error of VIS and SWIR images in the scanning direction was less than 1 pixel; these results proved that the system was stable.