Transfus Apher Sci ; 25(1): 5-15, 2001 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-11791762

RESUMO

On-line plasmapheresis using microporous membranes for filtration normally requires two external circulations (double treatment) before the detoxified plasma can be returned to the patient. The duomodule, a new filter arrangement developed by our group, integrates both steps in one equipment module using only one external circuit. Separations of aqueous polyethylene glycol (PEG) solutions as well as human plasma were carried out using the duomodule arrangement. The results revealed a considerable decrease of higher molecular substances in the feed solutions and a significant increase of these components in the permeate solution accumulated in the external compartment of the module. In conclusion, the duomodule arrangement seems to be an useful tool for the therapeutic apheresis.


Assuntos
Plasmaferese/instrumentação , Fracionamento Químico , Cromatografia Líquida , Desenho de Equipamento , Filtração/instrumentação , Humanos , Peso Molecular , Plasma , Polietilenoglicóis , Soluções