1.
J Appl Clin Med Phys ; 25(7): e14347, 2024 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-38576174

RESUMO

PURPOSE: This work investigated the dosimetric accuracy of the intensity-modulated bolus electron conformal therapy (IM-BECT) planning and delivery process using the decimal ElectronRT (eRT) treatment planning system. METHODS: An IM-BECT treatment plan was designed using eRT for a cylindrical, anthropomorphic retromolar trigone phantom. Treatment planning involved specification of beam parameters and design of a variable thickness wax bolus and Passive Radiotherapy Intensity Modulator for Electrons (PRIME) device, which was comprised of 33 tungsten island blocks of discrete diameters from 0.158 to 0.223 cm (Intensity Reduction Factors from 0.937 to 0.875, respectively) inside a 10.1 × 6.7 cm2 copper cutout. For comparison of calculation accuracy, a BECT plan was generated by copying the IM-BECT plan and removing the intensity modulation. For both plans, a 16 MeV electron beam was used with 104.7 cm source-to-surface distance to bolus. In-phantom TLD-100 measurements (N = 47) were compared with both eRT planned dose distributions, which used the pencil beam redefinition algorithm with modifications for passive electron intensity modulation (IM-PBRA). Dose difference and distance to agreement (DTA) metrics were computed for each measurement point. RESULTS: Comparison of measured dose distributions with planned dose distributions yielded dose differences (calculated minus measured) characterized by a mean and standard deviation of -0.36% ± 1.64% for the IM-BECT plan, which was similar to -0.36% ± 1.90% for the BECT plan. All dose measurements were within 5% of the planned dose distribution, with both the BECT and IM-BECT measurement sets having 46/47 (97.8%) points within 3% or within 3 mm of the respective treatment plans. CONCLUSIONS: It was found that the IM-BECT treatment plan generated using eRT was sufficiently accurate for clinical use when compared to TLD measurements in a cylindrical, anthropomorphic phantom, and was similarly accurate to the BECT treatment plan in the same phantom.


Assuntos
Elétrons , Imagens de Fantasmas , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador , Radioterapia de Intensidade Modulada , Humanos , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada/métodos , Elétrons/uso terapêutico , Algoritmos , Órgãos em Risco/efeitos da radiação , Radiometria/métodos , Cabeça/efeitos da radiação
2.
J Appl Clin Med Phys ; 24(7): e13943, 2023 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-36855930

RESUMO

PURPOSE: This work developed an x-ray-based method for performing factory quality assurance (QA) of Passive Radiotherapy Intensity Modulators for Electrons (PRIME) device fabrication. This method measures errors in position, diameter, and orientation of cylindrical island blocks on a hexagonal grid that comprises PRIME devices and the impact of such errors on the underlying intensity distribution. METHODS: X-ray images were acquired of six PRIME devices, which modeled three error cases (small random, large random, and systematic errors) for two island block diameters (0.158 and 0.352 cm). Island blocks in each device, 0.6 cm long tungsten cylinders of constant diameter, were spaced 0.6 cm on a hexagonal grid over approximately 8 cm square. Using a 50 kVp x-ray image, each island block projected a racetrack, whose perimeter was fit to a function that allowed determination of its position, diameter, and angular orientation (θ, ϕ). These measured parameters were input into a pencil beam algorithm (PBA) dose calculation performed in water (16 MeV, SSD = 103 cm) for each device. PBA calculated intensity distributions using measured and planned (exact) island block parameters were compared. RESULTS: Θ distributions for the 0.158 and 0.352 cm devices were nearly identical for each error case, with θ values for most island blocks being within 3.2°, 8.5°, and 7.5° for the small random, large random, and systematic error PRIME devices, respectively. Corresponding intensity differences between using measured and planned island block parameters were within 1.0% and 2.8% (small random), 2.2% and 4.8% (large random), and 3.2% and 6.7% (systematic) for the 0.158 and 0.352 cm devices, respectively. CONCLUSION: This approach provides a viable and economical method for factory QA of fabricated PRIME devices by determining errors in their planned intensity distribution from which their quality can be assessed prior to releasing to the customer.


Assuntos
Elétrons , Radioterapia de Intensidade Modulada , Humanos , Raios X , Radiografia , Planejamento da Radioterapia Assistida por Computador/métodos , Dosagem Radioterapêutica , Imagens de Fantasmas , Radioterapia de Intensidade Modulada/métodos
3.
J Appl Clin Med Phys ; 24(2): e13889, 2023 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-36610042

RESUMO

PURPOSE: Passive Radiotherapy Intensity Modulators for Electrons (PRIME) devices are comprised of cylindrical tungsten island blocks imbedded in a machinable foam slab within the patient's cutout. Intensity-modulated bolus electron conformal therapy (IM-BECT) uses PRIME devices to reduce dose heterogeneity caused by the irregular bolus surface. Heretofore, IM-BECT dose calculations used the pencil beam redefinition algorithm (PBRA) assuming perfect collimation. This study investigates modeling electron scatter into and out the sides of island blocks. METHODS: Dose distributions were measured in a water phantom at 7, 13, and 20 MeV for devices having nominal intensity reduction factors of 1.000 (foam only), 0.937, 0.812, and 0.688, corresponding to nominal island block diameters (dnom ) of 0.158, 0.273, and 0.352 cm, respectively. Pencil beam theory derived an effective diameter (dIS ) to account for in-scattered electrons as a function of dnom and beam energy (Ep,0 ). However, for out-scattered electrons, an effective diameter (dmod ) was estimated by best fitting measured data. RESULTS: In the modulated region (under island blocks, depth < R90 ), modified PBRA-calculated dose distributions showed 2%/2 mm passing rates for dnom  = 0.158, 0.273, and 0.352 cm of (100%, 100%, 100%) at 7 MeV, (100%, 100%, 93.5%) at 13 MeV, and (99.8%, 85.4%, and 71.5%) at 20 MeV. The largest dose differences (≤ 6%) occurred at the highest energy (20 MeV), largest dnom , shallowest depths (≤ 2 cm), and on central axis. CONCLUSIONS: An equation for modeling island block scatter, dmod (dnom , Ep,0 ), has been developed for use in the PBRA, insignificantly impacting calculation time. Although inaccuracy sometimes exceeded our 2%/2 mm criteria, it could be clinically acceptable, as superficial dose differences often fall inside the bolus. Also, patient PRIME devices are expected to have fewer large diameter island blocks than did test devices. Inaccuracies are attributed to out-scattered electrons having energy spectra different than the primary beams.


Assuntos
Elétrons , Radioterapia Conformacional , Humanos , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador , Imagens de Fantasmas
4.
J Appl Clin Med Phys ; 22(10): 8-21, 2021 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-34558774

RESUMO

PURPOSE: Bolus electron conformal therapy (BECT) is a clinically useful, well-documented, and available technology. The addition of intensity modulation (IM) to BECT reduces volumes of high dose and dose spread in the planning target volume (PTV). This paper demonstrates new techniques for a process that should be suitable for planning and delivering IM-BECT using passive radiotherapy intensity modulation for electrons (PRIME) devices. METHODS: The IM-BECT planning and delivery process is an addition to the BECT process that includes intensity modulator design, fabrication, and quality assurance. The intensity modulator (PRIME device) is a hexagonal matrix of small island blocks (tungsten pins of varying diameter) placed inside the patient beam-defining collimator (cutout). Its design process determines a desirable intensity-modulated electron beam during the planning process, then determines the island block configuration to deliver that intensity distribution (segmentation). The intensity modulator is fabricated and quality assurance performed at the factory (.decimal, LLC, Sanford, FL). Clinical quality assurance consists of measuring a fluence distribution in a plane perpendicular to the beam in a water or water-equivalent phantom. This IM-BECT process is described and demonstrated for two sites, postmastectomy chest wall and temple. Dose plans, intensity distributions, fabricated intensity modulators, and quality assurance results are presented. RESULTS: IM-BECT plans showed improved D90-10 over BECT plans, 6.4% versus 7.3% and 8.4% versus 11.0% for the postmastectomy chest wall and temple, respectively. Their intensity modulators utilized 61 (single diameter) and 246 (five diameters) tungsten pins, respectively. Dose comparisons for clinical quality assurance showed that for doses greater than 10%, measured agreed with calculated dose within 3% or 0.3 cm distance-to-agreement (DTA) for 99.9% and 100% of points, respectively. CONCLUSION: These results demonstrated the feasibility of translating IM-BECT to the clinic using the techniques presented for treatment planning, intensity modulator design and fabrication, and quality assurance processes.


Assuntos
Neoplasias da Mama , Radioterapia Conformacional , Elétrons , Feminino , Humanos , Mastectomia , Imagens de Fantasmas
5.
J Appl Clin Med Phys ; 21(12): 131-145, 2020 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-33207033

RESUMO

PURPOSE: This project determined the range of island block geometric configurations useful for the clinical utilization of intensity-modulated bolus electron conformal therapy (IM-BECT). METHODS: Multiple half-beam island block geometries were studied for seven electron energies 7-20 MeV at 100 and 103 cm source-to-surface distance (SSD). We studied relative fluence distributions at 0.5 cm and 2.0 cm depths in water, resulting in 28 unique beam conditions. For each beam condition, we studied intensity reduction factor (IRF) values of 0.70, 0.75, 0.80, 0.85, 0.90, and 0.95, and hexagonal packing separations for the island blocks of 0.50, 0.75, 1.00, 1.25, and 1.50 cm, that is, 30 unique IM configurations and 840 unique beam-IM combinations. A combination was deemed acceptable if the average intensity downstream of the intensity modulator agreed within 2% of that intended and the variation in fluence was less than ±2%. RESULTS: For 100 cm SSD, and for 0.5 cm depth, results showed that beam energies above 13 MeV did not exhibit sufficient scatter to produce clinically acceptable fluence (intensity) distributions for all IRF values (0.70-0.95). In particular, 20 MeV fluence distributions were unacceptable for any values, and acceptable 16 MeV fluence distributions were limited to a minimum IRF of 0.85. For the 2.0 cm depth, beam energies up to and including 20 MeV had acceptable fluence distributions. For 103 cm SSD and for 0.5 cm and 2.0 cm depths, results showed that all beam energies (7-20 MeV) had clinically acceptable fluence distributions for all IRF values (0.70-0.95). In general, the more clinically likely 103 cm SSD had acceptable fluence distributions with larger separations (r), which allow larger block diameters. CONCLUSION: The geometric operating range of island block separations and IRF values (block diameters) producing clinically appropriate IM electron beams has been determined.


Assuntos
Elétrons , Radioterapia Conformacional , Humanos , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador
6.
J Appl Clin Med Phys ; 19(4): 75-86, 2018 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-29756267

RESUMO

PURPOSE: This study evaluated a new electron collimation system design for Elekta 6-20 MeV beams, which should reduce applicator weights by 25%-30%. Such reductions, as great as 3.9 kg for the largest applicator, should result in considerably easier handling by members of the radiotherapy team. METHODS: Prototype 10 × 10 and 20 × 20-cm2 applicators, used to measure weight, in-field flatness, and out-of-field leakage dose, were constructed according to the previously published design with two minor modifications: (a) rather than tungsten, lead was used for trimmer material; and (b) continuous trimmer outer-edge bevel was approximated by three steps. Because of lead plate softness, a 0.32-cm aluminum plate replaced the equivalent lead thickness on the trimmer's downstream surface for structural support. Models of all applicators (6 × 6-25 × 25 cm2 ) with these modifications were inserted into a Monte Carlo (MC) model for dose calculations using 7, 13, and 20 MeV beams. Planar dose distributions were measured and calculated at 1- and 2-cm water depths to evaluate in-field beam flatness and out-of-field leakage dose. RESULTS: Prototype 10 × 10 and 20 × 20-cm2 applicator measurements agreed with calculated weights, in-field flatness, and out-of-field leakage doses for 7, 13, and 20 MeV beams. Also, MC dose calculations showed that all applicators (6 × 6-25 × 25 cm2 ) and 7, 13, and 20 MeV beams met our stringent in-field flatness specifications (±3% major axes; ±4% diagonals) and IEC out-of-field leakage dose specifications. CONCLUSIONS: Our results validated the new electron collimating system design for Elekta 6-20 MeV electron beams, which could serve as basis for a new clinical electron collimating system with significantly reduced applicator weights.


Assuntos
Aceleradores de Partículas , Elétrons , Método de Monte Carlo , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador
7.
J Appl Clin Med Phys ; 19(3): 183-192, 2018 May.
Artigo em Inglês | MEDLINE | ID: mdl-29603874

RESUMO

This study compares energy spectra of the multiple electron beams of individual radiotherapy machines, as well as the sets of spectra across multiple matched machines. Also, energy spectrum metrics are compared with central-axis percent depth-dose (PDD) metrics. METHODS: A lightweight, permanent magnet spectrometer was used to measure energy spectra for seven electron beams (7-20 MeV) on six matched Elekta Infinity accelerators with the MLCi2 treatment head. PDD measurements in the distal falloff region provided R50 and R80-20 metrics in Plastic Water® , which correlated with energy spectrum metrics, peak mean energy (PME) and full-width at half maximum (FWHM). RESULTS: Visual inspection of energy spectra and their metrics showed whether beams on single machines were properly tuned, i.e., FWHM is expected to increase and peak height decrease monotonically with increased PME. Also, PME spacings are expected to be approximately equal for 7-13 MeV beams (0.5-cm R90 spacing) and for 13-16 MeV beams (1.0-cm R90 spacing). Most machines failed these expectations, presumably due to tolerances for initial beam matching (0.05 cm in R90 ; 0.10 cm in R80-20 ) and ongoing quality assurance (0.2 cm in R50 ). Also, comparison of energy spectra or metrics for a single beam energy (six machines) showed outlying spectra. These variations in energy spectra provided ample data spread for correlating PME and FWHM with PDD metrics. Least-squares fits showed that R50 and R80-20 varied linearly and supralinearly with PME, respectively; however, both suggested a secondary dependence on FWHM. Hence, PME and FWHM could serve as surrogates for R50 and R80-20 for beam tuning by the accelerator engineer, possibly being more sensitive (e.g., 0.1 cm in R80-20 corresponded to 2.0 MeV in FWHM). CONCLUSIONS: Results of this study suggest a lightweight, permanent magnet spectrometer could be a useful beam-tuning instrument for the accelerator engineer to (a) match electron beams prior to beam commissioning, (b) tune electron beams for the duration of their clinical use, and (c) provide estimates of PDD metrics following machine maintenance. However, a real-time version of the spectrometer is needed to be practical.


Assuntos
Elétrons , Método de Monte Carlo , Neoplasias/radioterapia , Aceleradores de Partículas/instrumentação , Imagens de Fantasmas , Planejamento da Radioterapia Assistida por Computador/métodos , Humanos , Dosagem Radioterapêutica , Radioterapia de Intensidade Modulada/métodos
8.
J Appl Clin Med Phys ; 18(6): 10-19, 2017 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-28875590

RESUMO

This work introduces a new technology for electron intensity modulation, which uses small area island blocks within the collimating aperture and small area island apertures in the collimating insert. Due to multiple Coulomb scattering, electrons contribute dose under island blocks and lateral to island apertures. By selecting appropriate lateral positions and diameters of a set of island blocks and island apertures, for example, a hexagonal grid with variable diameter circular island blocks, intensity modulated beams can be produced for appropriate air gaps between the intensity modulator (position of collimating insert) and the patient. Such a passive radiotherapy intensity modulator for electrons (PRIME) is analogous to using physical attenuators (metal compensators) for intensity modulated x-ray therapy (IMXT). For hexagonal spacing, the relationship between block (aperture) separation (r) and diameter (d) and the local intensity reduction factor (IRF) is discussed. The PRIME principle is illustrated using pencil beam calculations for select beam geometries in water with half beams modulated by 70%-95% and for one head and neck field of a patient treated with bolus electron conformal therapy. Proof of principle is further illustrated by showing agreement between measurement and calculation for a prototype PRIME. Potential utilization of PRIME for bolus electron conformal therapy, segmented-field electron conformal therapy, modulated electron radiation therapy, and variable surface geometries is discussed. Further research and development of technology for the various applications is discussed. In summary, this paper introduces a practical, new technology for electron intensity modulation in the clinic, demonstrates proof of principle, discusses potential clinical applications, and suggests areas of further research and development.


Assuntos
Elétrons/uso terapêutico , Radioterapia de Intensidade Modulada/métodos , Humanos
9.
J Appl Clin Med Phys ; 18(5): 259-270, 2017 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-28801965

RESUMO

Prototype 10 × 10 and 20 × 20-cm2 electron collimators were designed for the Elekta Infinity accelerator (MLCi2 treatment head), with the goal of reducing the trimmer weight of excessively heavy current applicators while maintaining acceptable beam flatness (±3% major axes, ±4% diagonals) and IEC leakage dose. Prototype applicators were designed initially using tungsten trimmers of constant thickness (1% electron transmission) and cross-sections with inner and outer edges positioned at 95% and 2% off-axis ratios (OARs), respectively, cast by the upstream collimating component. Despite redefining applicator size at isocenter (not 5 cm upstream) and reducing the energy range from 4-22 to 6-20 MeV, the designed 10 × 10 and 20 × 20-cm2 applicator trimmers weighed 6.87 and 10.49 kg, respectively, exceeding that of the current applicators (5.52 and 8.36 kg, respectively). Subsequently, five design modifications using analytical and/or Monte Carlo (MC) calculations were applied, reducing trimmer weight while maintaining acceptable in-field flatness and mean leakage dose. Design Modification 1 beveled the outer trimmer edges, taking advantage of only low-energy beams scattering primary electrons sufficiently to reach the outer trimmer edge. Design Modification 2 optimized the upper and middle trimmer distances from isocenter for minimal trimmer weights. Design Modification 3 moved inner trimmer edges inward, reducing trimmer weight. Design Modification 4 determined optimal X-ray jaw positions for each energy. Design Modification 5 adjusted middle and lower trimmer shapes and reduced upper trimmer thickness by 50%. Design Modifications 1→5 reduced trimmer weights from 6.87→5.86→5.52→5.87→5.43→3.73 kg for the 10 × 10-cm2 applicator and 10.49→9.04→8.62→7.73→7.35→5.09 kg for the 20 × 20-cm2 applicator. MC simulations confirmed these final designs produced acceptable in-field flatness and met IEC-specified leakage dose at 7, 13, and 20 MeV. These results allowed collimation system design for 6 × 6-25 × 25-cm2 applicators. Reducing trimmer weights by as much as 4 kg (25 × 25-cm2 applicator) should result in easier applicator handling by the radiotherapy team.


Assuntos
Elétrons/uso terapêutico , Aceleradores de Partículas , Desenho de Equipamento , Humanos , Método de Monte Carlo , Dosagem Radioterapêutica
10.
J Appl Clin Med Phys ; 17(5): 157-176, 2016 09 08.
Artigo em Inglês | MEDLINE | ID: mdl-27685101

RESUMO

This study provided baseline data required for a greater project, whose objective was to design a new Elekta electron collimation system having significantly lighter electron applicators with equally low out-of field leakage dose. Specifically, off-axis dose profiles for the electron collimation system of our uniquely configured Elekta Infinity accelerator with the MLCi2 treatment head were measured and calculated for two primary purposes: 1) to evaluate and document the out-of-field leakage dose in the patient plane and 2) to validate the dose distributions calculated using a BEAMnrc Monte Carlo (MC) model for out-of-field dose profiles. Off-axis dose profiles were measured in a water phantom at 100 cm SSD for 1 and 2 cm depths along the in-plane, cross-plane, and both diagonal axes using a cylindrical ionization chamber with the 10 × 10 and 20 × 20 cm2 applicators and 7, 13, and 20 MeV beams. Dose distributions were calculated using a previously developed BEAMnrc MC model of the Elekta Infinity accelerator for the same beam energies and applicator sizes and compared with measurements. Measured results showed that the in-field beam flatness met our acceptance criteria (± 3% on major and ±4% on diagonal axes) and that out-of-field mean and maximum percent leakage doses in the patient plane met acceptance criteria as specified by the International Electrotechnical Commission (IEC). Cross-plane out-of-field dose profiles showed greater leakage dose than in-plane profiles, attributed to the curved edges of the upper X-ray jaws and multileaf collimator. Mean leakage doses increased with beam energy, being 0.93% and 0.85% of maximum central axis dose for the 10 × 10 and 20 × 20 cm2 applicators, respectively, at 20 MeV. MC calculations predicted the measured dose to within 0.1% in most profiles outside the radiation field; however, excluding model-ing of nontrimmer applicator components led to calculations exceeding measured data by as much as 0.2% for some regions along the in-plane axis. Using EGSnrc LATCH bit filtering to separately calculate out-of-field leakage dose components (photon dose, primary electron dose, and electron dose arising from interactions in various collimating components), MC calculations revealed that the primary electron dose in the out-of-field leakage region was small and decreased as beam energy increased. Also, both the photon dose component and electron dose com-ponent resulting from collimator scatter dominated the leakage dose, increasing with increasing beam energy. We concluded that our custom Elekta Infinity with the MLCi2 treatment head met IEC leakage dose criteria in the patient plane. Also, accuracy of our MC model should be sufficient for our use in the design of a new, improved electron collimation system.


Assuntos
Elétrons , Cabeça , Aceleradores de Partículas/instrumentação , Humanos , Processamento de Imagem Assistida por Computador/métodos , Método de Monte Carlo , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Espalhamento de Radiação
11.
J Appl Clin Med Phys ; 17(5): 245­261, 2016 09 08.
Artigo em Inglês | MEDLINE | ID: mdl-27685126

RESUMO

The purpose of this work was to evaluate differences in dose resulting from the use of copper aperture inserts compared to lead-alloy (Cerrobend) aperture inserts for electron beam therapy. Specifically, this study examines if copper aperture inserts can be used clinically with the same commissioning data measured using lead-alloy aperture inserts. The copper inserts were acquired from .decimal, LLC and matching lead-alloy, Cerrobend inserts were constructed in-house for 32 com-binations of nine square insert field sizes (2 × 2 to 20 × 20 cm2) and five applicator sizes (6 × 6 to 25 × 25 cm2). Percent depth-dose and off-axis relative dose profiles were measured using an electron diode in water for select copper and Cerrobend inserts for a subset of applicators (6 × 6, 10 × 10, 25 × 25 cm2) and energies (6, 12, 20 MeV) at 100 and 110 cm source-to-surface distances (SSD) on a Varian Clinac 21EX accelerator. Dose outputs were measured for all field size-insert combina-tions and five available energies (6-20 MeV) at 100 cm SSD and for a smaller subset at 110 cm SSD. Using these data, 2D planar absolute dose distributions were generated and compared. Criteria for agreement were ± 2% of maximum dose or 1 mm distance-to-agreement for 99% of points. A gamma analysis of the beam dosimetry showed 94 of 96 combinations of insert size, applicator, energy, and SSD were within the 2%/1 mm criteria for > 99% of points. Outside the field, copper inserts showed less bremsstrahlung dose under the insert compared to Cerrobend (greatest difference was 2.5% at 20 MeV and 100 cm SSD). This effect was most prominent at the highest energies for combinations of large applicators with small field sizes, causing some gamma analysis failures. Inside the field, more electrons scattered from the collimator edge of copper compared to Cerrobend, resulting in an increased dose at the field edge for copper at shallow depths (greatest increase was 1% at 20 MeV and 100 cm SSD). Dose differences decreased as the SSD increased, with no gamma failures at 110 cm SSD. Inserts for field sizes ≥ 6 × 6 cm2 at any energy, or for small fields (≤ 4 × 4 cm2) at energies < 20 MeV, showed dosimetric differences less than 2%/1 mm for more than 99% of points. All areas of comparison criteria failures were from lower out-of-field dose under copper inserts due to a reduction in bremsstrahlung production, which is clinically beneficial in reducing dose to healthy tissue outside of the planned treatment volume. All field size-applicator size-energy combinations passed 3%/1 mm criteria for 100% of points. Therefore, it should be clinically acceptable to utilize copper insets with dose distributions measured with Cerrobend inserts for treatment planning dose calculations and monitor unit calculations.


Assuntos
Ligas/química , Cobre/química , Elétrons , Chumbo/química , Imagens de Fantasmas , Humanos , Método de Monte Carlo , Aceleradores de Partículas , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada/métodos
12.
J Appl Clin Med Phys ; 17(3): 52-60, 2016 05 08.
Artigo em Inglês | MEDLINE | ID: mdl-27167259

RESUMO

The purpose of this study was to evaluate the accuracy and calculation speed of electron dose distributions calculated by the Eclipse electron Monte Carlo (eMC) algorithm for use with bolus electron conformal therapy (ECT). The recent com-mercial availability of bolus ECT technology requires further validation of the eMC dose calculation algorithm. eMC-calculated electron dose distributions for bolus ECT have been compared to previously measured TLD-dose points throughout patient-based cylindrical phantoms (retromolar trigone and nose), whose axial cross sections were based on the mid-PTV (planning treatment volume) CT anatomy. The phantoms consisted of SR4 muscle substitute, SR4 bone substitute, and air. The treatment plans were imported into the Eclipse treatment planning system, and electron dose distributions calculated using 1% and < 0.2% statistical uncertainties. The accuracy of the dose calculations using moderate smoothing and no smooth-ing were evaluated. Dose differences (eMC-calculated less measured dose) were evaluated in terms of absolute dose difference, where 100% equals the given dose, as well as distance to agreement (DTA). Dose calculations were also evaluated for calculation speed. Results from the eMC for the retromolar trigone phantom using 1% statistical uncertainty without smoothing showed calculated dose at 89% (41/46) of the measured TLD-dose points was within 3% dose difference or 3 mm DTA of the measured value. The average dose difference was -0.21%, and the net standard deviation was 2.32%. Differences as large as 3.7% occurred immediately distal to the mandible bone. Results for the nose phantom, using 1% statistical uncertainty without smoothing, showed calculated dose at 93% (53/57) of the measured TLD-dose points within 3% dose difference or 3 mm DTA. The average dose difference was 1.08%, and the net standard deviation was 3.17%. Differences as large as 10% occurred lateral to the nasal air cavities. Including smoothing had insignificant effects on the accuracy of the retromolar trigone phantom calculations, but reduced the accuracy of the nose phantom calculations in the high-gradient dose areas. Dose calculation times with 1% statistical uncertainty for the retromolar trigone and nose treatment plans were 30 s and 24 s, respectively, using 16 processors (Intel Xeon E5-2690, 2.9 GHz) on a framework agent server (FAS). In comparison, the eMC was significantly more accurate than the pencil beam algorithm (PBA). The eMC has comparable accuracy to the pencil beam redefinition algorithm (PBRA) used for bolus ECT planning and has acceptably low dose calculation times. The eMC accuracy decreased when smoothing was used in high-gradient dose regions. The eMC accuracy was consistent with that previously reported for accuracy of the eMC electron dose algorithm and shows that the algorithm is suitable for clinical implementation of bolus ECT.


Assuntos
Algoritmos , Elétrons , Método de Monte Carlo , Imagens de Fantasmas , Planejamento da Radioterapia Assistida por Computador/normas , Radioterapia Conformacional/normas , Humanos , Dosagem Radioterapêutica
13.
Med Phys ; 42(9): 5517-29, 2015 Sep.
Artigo em Inglês | MEDLINE | ID: mdl-26328999

RESUMO

PURPOSE: The purpose of this work was to adapt a lightweight, permanent magnet electron energy spectrometer for the measurement of energy spectra of therapeutic electron beams. METHODS: An irradiation geometry and measurement technique were developed for an approximately 0.54-T, permanent dipole magnet spectrometer to produce suitable latent images on computed radiography (CR) phosphor strips. Dual-pinhole electron collimators created a 0.318-cm diameter, approximately parallel beam incident on the spectrometer and an appropriate dose rate at the image plane (CR strip location). X-ray background in the latent image, reduced by a 7.62-cm thick lead block between the pinhole collimators, was removed using a fitting technique. Theoretical energy-dependent detector response functions (DRFs) were used in an iterative technique to transform CR strip net mean dose profiles into energy spectra on central axis at the entrance to the spectrometer. These spectra were transformed to spectra at 95-cm source to collimator distance (SCD) by correcting for the energy dependence of electron scatter. The spectrometer was calibrated by comparing peak mean positions in the net mean dose profiles, initially to peak mean energies determined from the practical range of central-axis percent depth-dose (%DD) curves, and then to peak mean energies that accounted for how the collimation modified the energy spectra (recalibration). The utility of the spectrometer was demonstrated by measuring the energy spectra for the seven electron beams (7-20 MeV) of an Elekta Infinity radiotherapy accelerator. RESULTS: Plots of DRF illustrated their dependence on energy and position in the imaging plane. Approximately 15 iterations solved for the energy spectra at the spectrometer entrance from the measured net mean dose profiles. Transforming those spectra into ones at 95-cm SCD increased the low energy tail of the spectra, while correspondingly decreasing the peaks and shifting them to slightly lower energies. Energy calibration plots of peak mean energy versus peak mean position of the net mean dose profiles for each of the seven electron beams followed the shape predicted by the Lorentz force law for a uniform z-component of the magnetic field, validating its being modeled as uniform (0.542 ± 0.027 T). Measured Elekta energy spectra and their peak mean energies correlated with the 0.5-cm (7-13 MeV) and the 1.0-cm (13-20 MeV) R90 spacings of the %DD curves. The full-width-half-maximum of the energy spectra decreased with decreasing peak mean energy with the exception of the 9-MeV beam, which was anomalously wide. Similarly, R80-20 decreased linearly with peak mean energy with the exception of the 9 MeV beam. Both were attributed to suboptimal tuning of the high power phase shifter for the recycled radiofrequency power reentering the traveling wave accelerator. CONCLUSIONS: The apparatus and analysis techniques of the authors demonstrated that an inexpensive, lightweight, permanent magnet electron energy spectrometer can be used for measuring the electron energy distributions of therapeutic electron beams (6-20 MeV). The primary goal of future work is to develop a real-time spectrometer by incorporating a real-time imager, which has potential applications such as beam matching, ongoing beam tune maintenance, and measuring spectra for input into Monte Carlo beam calculations.


Assuntos
Elétrons/uso terapêutico , Imãs , Aceleradores de Partículas , Radioterapia/instrumentação , Análise Espectral/instrumentação
14.
J Appl Clin Med Phys ; 15(6): 4849, 2014 Nov 08.
Artigo em Inglês | MEDLINE | ID: mdl-25493509

RESUMO

The purpose of this work was to develop a user friendly, accurate, real-time com- puter simulator to facilitate the design of dual foil scattering systems for electron beams on radiotherapy accelerators. The simulator allows for a relatively quick, initial design that can be refined and verified with subsequent Monte Carlo (MC) calculations and measurements. The simulator also is a powerful educational tool. The simulator consists of an analytical algorithm for calculating electron fluence and X-ray dose and a graphical user interface (GUI) C++ program. The algorithm predicts electron fluence using Fermi-Eyges multiple Coulomb scattering theory with the reduced Gaussian formalism for scattering powers. The simulator also estimates central-axis and off-axis X-ray dose arising from the dual foil system. Once the geometry of the accelerator is specified, the simulator allows the user to continuously vary primary scattering foil material and thickness, secondary scat- tering foil material and Gaussian shape (thickness and sigma), and beam energy. The off-axis electron relative fluence or total dose profile and central-axis X-ray dose contamination are computed and displayed in real time. The simulator was validated by comparison of off-axis electron relative fluence and X-ray percent dose profiles with those calculated using EGSnrc MC. Over the energy range 7-20 MeV, using present foils on an Elekta radiotherapy accelerator, the simulator was able to reproduce MC profiles to within 2% out to 20 cm from the central axis. The central-axis X-ray percent dose predictions matched measured data to within 0.5%. The calculation time was approximately 100 ms using a single Intel 2.93 GHz processor, which allows for real-time variation of foil geometrical parameters using slider bars. This work demonstrates how the user-friendly GUI and real-time nature of the simulator make it an effective educational tool for gaining a better understanding of the effects that various system parameters have on a relative dose profile. This work also demonstrates a method for using the simulator as a design tool for creating custom dual scattering foil systems in the clinical range of beam energies (6-20 MeV). 


Assuntos
Simulação por Computador , Planejamento da Radioterapia Assistida por Computador/instrumentação , Planejamento da Radioterapia Assistida por Computador/métodos , Espalhamento de Radiação , Algoritmos , Elétrons , Humanos , Método de Monte Carlo , Aceleradores de Partículas , Dosagem Radioterapêutica , Raios X
15.
Radiat Res ; 182(6): 607-17, 2014 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-25409122

RESUMO

The goal of the current study was to measure the energy dependence of survival of rat 9L glioma cells labeled with iododeoxyuridine (IUdR) that underwent photon-activated Auger electron therapy using 25-35 keV monochromatic X rays, i.e., above and below the K-edge energy of iodine. Rat 9L glioma cells were selected because of their radioresistance, ability to be implanted for future in vivo studies and analogy to radioresistant human gliomas. Survival curves were measured for a 4 MV X-ray beam and synchrotron produced monochromatic 35, 30 and 25 keV X-ray beams. IUdR was incorporated into the DNA at levels of 0, 9 and 18% thymidine replacement for 4 MV and 35 keV and 0 and 18% thymidine replacement for 30 and 25 keV. For 10 combinations of beam energy and thymidine replacement, 62 data sets (3-13 per combination) provided 776 data points (47-148 per combination). Survival versus dose data taken for the same combination, but on different days, were merged by including the zero-dose points in the nonlinear, chi-squared data fitting using the linear-quadratic model and letting the best estimate to the zero-dose plating efficiency for each of the different days be a fitting parameter. When comparing two survival curves, the ratio of doses resulting in 10% survival gave sensitization enhancement ratios (SER10) from which contributions due to linear energy transfer (LET) (SER10,LET), IUdR radiosensitization (SER10,RS), the Auger effect (SER10,AE) and the total of all effects (SER10,T) were determined. At 4 MV and 35, 30 and 25 keV, SER10,LET values were 1.00, 1.08 ± 0.03, 1.22 ± 0.02 and 1.37 ± 0.02, respectively. At 4 MV SER10,RS values for 9 and 18% IUdR were 1.28 ± 0.02 and 1.40 ± 0.02, respectively. Assuming LET effects were independent of percentage IUdR and radiosensitization effects were independent of energy, SER10,AE values for 18% IUdR at 35, 30 and 25 keV were 1.35 ± 0.05, 1.06 ± 0.03 and 0.98 ± 0.03, respectively. The value for 9% IUdR at 35 keV was 1.01 ± 0.04. First, we found the radioresistant rat 9L glioma cell line exhibited an SER10 due to the Auger effect of 1.35 at (35 keV, 18% IUdR) and an SER10 due to the radiosensitizing effect of 1.40 at (4 MV, 18% IUdR), both significantly less than values for previously reported cell lines. These low individual values emphasize the benefit of their combined value (SER10 of approximately 1.9) for achieving clinical benefit. Second, as expected, we observed that energies below the K-edge of iodine (25 and 30 keV), for which there are L, M and higher shell photoelectric events creating Auger electrons, show no promise for Auger electron therapy. Third, to proceed with future in vivo studies, additional data from 35-65 keV are needed to determine the optimal X-ray energy for IUdR Auger electron therapy. Only then can there be an answer to the question, how well the energy dependence of in vitro survival data supports the potential for photon-activated Auger electron therapy with IUdR in cancer radiotherapy.


Assuntos
Elétrons/uso terapêutico , Glioma/patologia , Idoxuridina/farmacologia , Fótons/uso terapêutico , Radiossensibilizantes/farmacologia , Animais , Linhagem Celular Tumoral , Sobrevivência Celular/efeitos dos fármacos , Sobrevivência Celular/efeitos da radiação , Humanos , Transferência Linear de Energia , Ratos
16.
J Appl Clin Med Phys ; 15(2): 4490, 2014 Mar 06.
Artigo em Inglês | MEDLINE | ID: mdl-24710434

RESUMO

In 2009, Mary Bird Perkins Cancer Center (MBPCC) established a Radiation Oncology Physics Residency Program to provide opportunities for medical physics residency training to MS and PhD graduates of the CAMPEP-accredited Louisiana State University (LSU)-MBPCC Medical Physics Graduate Program. The LSU-MBPCC Program graduates approximately six students yearly, which equates to a need for up to twelve residency positions in a two-year program. To address this need for residency positions, MBPCC has expanded its Program by developing a Consortium consisting of partnerships with medical physics groups located at other nearby clinical institutions. The consortium model offers the residents exposure to a broader range of procedures, technology, and faculty than available at the individual institutions. The Consortium institutions have shown a great deal of support from their medical physics groups and administrations in developing these partnerships. Details of these partnerships are specified within affiliation agreements between MBPCC and each participating institution. All partner sites began resident training in 2011. The Consortium is a network of for-profit, nonprofit, academic, community, and private entities. We feel that these types of collaborative endeavors will be required nationally to reach the number of residency positions needed to meet the 2014 ABR certification requirements and to maintain graduate medical physics training programs.


Assuntos
Certificação , Física Médica/educação , Internato e Residência/normas , Radioterapia (Especialidade)/educação , Humanos , Médicos , Desenvolvimento de Programas , Avaliação de Programas e Projetos de Saúde , Sociedades Médicas , Estados Unidos
17.
Pract Radiat Oncol ; 4(1): e67-73, 2014.
Artigo em Inglês | MEDLINE | ID: mdl-24621434

RESUMO

PURPOSE: To assess the accuracy and precision of cone-beam computed tomography (CBCT)-guided intensity modulated radiation therapy (IMRT). METHODS AND MATERIALS: A 7-field intensity modulated radiation therapy plan was constructed for an anthropomorphic head phantom loaded with a custom cassette containing radiochromic film. The phantom was positioned on the treatment table at 9 locations: 1 "correct" position and 8 "misaligned" positions along 3 orthogonal axes. A commercial kilovoltage cone-beam computed tomography (kV-CBCT) system (VolumeView, Elekta AB, Stockholm, Sweden) was then used to align the phantom prior to plan delivery. The treatment plan was delivered using the radiation therapy delivery system (Infinity; Elekta AB) 3 times for each of the 9 positions, allowing film measurement of the delivered dose distribution in 3 orthogonal planes. Comparison of the planned and delivered dose profiles along the major axes provided an estimate of the accuracy and precision of CBCT-guided IMRT. RESULTS: On average, targeting accuracy was found to be within 1 mm in all 3 major anatomic planes. Over all 54 measured dose profiles, the means and standard errors of the displacement of the center of the field between the measured and calculated profiles for each of the right-left, anterior-posterior, and superior-inferior axes were +0.08 ± 0.07 mm, +0.60 ± 0.08 mm, and +0.78 ± 0.16 mm, respectively. Agreement between planned and measured 80% profiles was less than 0.4 mm on either side along the right-left axis. A systematic shift of the measured profile of slightly less than 1 mm in anterior and superior directions was noted along the anterior-posterior and superior-inferior axes, respectively. CONCLUSIONS: Submillimeter targeting accuracy can be achieved using a commercial kV-CBCT IGRT system.


Assuntos
Tomografia Computadorizada de Feixe Cônico/métodos , Planejamento da Radioterapia Assistida por Computador/métodos , Radioterapia de Intensidade Modulada/métodos , Encéfalo/anatomia & histologia , Tomografia Computadorizada de Feixe Cônico/normas , Precisão da Medição Dimensional , Cabeça/anatomia & histologia , Humanos , Imagens de Fantasmas , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador/normas , Radioterapia de Intensidade Modulada/normas
18.
Med Phys ; 40(7): 071720, 2013 Jul.
Artigo em Inglês | MEDLINE | ID: mdl-23822424

RESUMO

PURPOSE: The purpose of this study was to document the improved accuracy of the pencil beam redefinition algorithm (PBRA) compared to the pencil beam algorithm (PBA) for bolus electron conformal therapy using cylindrical patient phantoms based on patient computed tomography (CT) scans of retromolar trigone and nose cancer. METHODS: PBRA and PBA electron dose calculations were compared with measured dose in retromolar trigone and nose phantoms both with and without bolus. For the bolus treatment plans, a radiation oncologist outlined a planning target volume (PTV) on the central axis slice of the CT scan for each phantom. A bolus was designed using the planning.decimal(®) (p.d) software (.decimal, Inc., Sanford, FL) to conform the 90% dose line to the distal surface of the PTV. Dose measurements were taken with thermoluminescent dosimeters placed into predrilled holes. The Pinnacle(3) (Philips Healthcare, Andover, MD) treatment planning system was used to calculate PBA dose distributions. The PBRA dose distributions were calculated with an in-house C++ program. In order to accurately account for the phantom materials a table correlating CT number to relative electron stopping and scattering powers was compiled and used for both PBA and PBRA dose calculations. Accuracy was determined by comparing differences in measured and calculated dose, as well as distance to agreement for each measurement point. RESULTS: The measured doses had an average precision of 0.9%. For the retromolar trigone phantom, the PBRA dose calculations had an average ± 1σ dose difference (calculated - measured) of -0.65% ± 1.62% without the bolus and -0.20% ± 1.54% with the bolus. The PBA dose calculation had an average dose difference of 0.19% ± 3.27% without the bolus and -0.05% ± 3.14% with the bolus. For the nose phantom, the PBRA dose calculations had an average dose difference of 0.50% ± 3.06% without bolus and -0.18% ± 1.22% with the bolus. The PBA dose calculations had an average dose difference of 0.65% ± 6.21% without bolus and 1.75% ± 5.94% with the bolus. From a clinical perspective an agreement of 5% or better between planned (calculated) and delivered (measured) dose is desired. Statistically, this was true for 99% (± 2σ) of the dose points for three of the four cases for the PBRA dose calculations, the exception being the nose without bolus for which this was true for 89% (± 1.6σ) of the dose points. For the retromolar trigone, with and without bolus, the PBA showed agreement of 5% or better for approximately 86% (± 1.5σ) of the dose points. For the nose, with and without bolus, the PBA showed agreement of 5% or better for only approximately 58% (± 0.8σ) of the dose points. CONCLUSIONS: The measured data, whose high precision makes them useful for evaluation of the accuracy of electron dose algorithms, will be made publicly available. Based on the spread in dose differences, the PBRA has at least twice the accuracy of the PBA. From a clinical perspective the PBRA accuracy is acceptable in the retromolar trigone and nose for electron therapy with and without bolus.


Assuntos
Algoritmos , Elétrons/uso terapêutico , Imagens de Fantasmas , Doses de Radiação , Radioterapia Conformacional/instrumentação , Humanos , Dosagem Radioterapêutica , Dosimetria Termoluminescente
19.
Med Phys ; 40(2): 021724, 2013 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-23387747

RESUMO

PURPOSE: The purpose of this study was to demonstrate that a bolus electron conformal therapy (ECT) dose plan and a mixed beam plan, composed of an intensity modulated x-ray therapy (IMXT) dose plan optimized on top of the bolus ECT plan, can be accurately delivered. METHODS: Calculated dose distributions were compared with measured dose distributions for parotid and chest wall (CW) bolus ECT and mixed beam plans, each simulated in a cylindrical polystyrene phantom that allowed film dose measurements. Bolus ECT plans were created for both parotid and CW PTVs (planning target volumes) using 20 and 16 MeV beams, respectively, whose 90% dose surface conformed to the PTV. Mixed beam plans consisted of an IMXT dose plan optimized on top of the bolus ECT dose plan. The bolus ECT, IMXT, and mixed beam dose distributions were measured using radiographic films in five transverse and one sagittal planes for a total of 36 measurement conditions. Corrections for film dose response, effects of edge-on photon irradiation, and effects of irregular phantom optical properties on the Cerenkov component of the film signal resulted in high precision measurements. Data set consistency was verified by agreement of depth dose at the intersections of the sagittal plane with the five measured transverse planes. For these same depth doses, results for the mixed beam plan agreed with the sum of the individual depth doses for the bolus ECT and IMXT plans. The six mean measured planar dose distributions were compared with those calculated by the treatment planning system for all modalities. Dose agreement was assessed using the 4% dose difference and 0.2 cm distance to agreement. RESULTS: For the combined high-dose region and low-dose region, pass rates for the parotid and CW plans were 98.7% and 96.2%, respectively, for the bolus ECT plans and 97.9% and 97.4%, respectively, for the mixed beam plans. For the high-dose gradient region, pass rates for the parotid and CW plans were 93.1% and 94.62%, respectively, for the bolus ECT plans and 89.2% and 95.1%, respectively, for the mixed beam plans. For all regions, pass rates for the parotid and CW plans were 98.8% and 97.3%, respectively, for the bolus ECT plans and 97.5% and 95.9%, respectively, for the mixed beam plans. For the IMXT component of the mixed beam plans, pass rates for the parotid and CW plans were 93.7% and 95.8%. CONCLUSIONS: Bolus ECT and mixed beam therapy dose delivery to the phantom were more accurate than IMXT delivery, adding confidence to the use of planning, fabrication, and delivery for bolus ECT tools either alone or as part of mixed beam therapy. The methodology reported in this work could serve as a basis for future standardization of the commissioning of bolus ECT or mixed beam therapy. When applying this technology to patients, it is recommended that an electron dose algorithm more accurate than the pencil beam algorithm, e.g., a Monte Carlo algorithm or analytical transport such as the pencil beam redefinition algorithm, be used for planning to ensure the desired accuracy.


Assuntos
Elétrons/uso terapêutico , Radioterapia de Intensidade Modulada/métodos , Humanos , Radiometria , Dosagem Radioterapêutica , Planejamento da Radioterapia Assistida por Computador
20.
Med Phys ; 39(12): 7412-7, 2012 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-23231291

RESUMO

PURPOSE: This work investigates the dose-response curves of GAFCHROMIC(®) EBT, EBT2, and EBT3 radiochromic films using synchrotron-produced monochromatic x-ray beams. EBT2 film is being utilized for dose verification in photoactivated Auger electron therapy at the Louisiana State University Center for Advanced Microstructures and Devices (CAMD) synchrotron facility. METHODS: Monochromatic beams of 25, 30, and 35 keV were generated on the tomography beamline at CAMD. Ion chamber depth-dose measurements were used to determine the dose delivered to films irradiated at depths from 0.7 to 8.5 cm in a 10 × 10 × 10-cm(3) polymethylmethacrylate phantom. AAPM TG-61 protocol was applied to convert measured ionization into dose. Films were digitized using an Epson 1680 Professional flatbed scanner and analyzed using the net optical density (NOD) derived from the red channel. A dose-response curve was obtained at 35 keV for EBT film, and at 25, 30, and 35 keV for EBT2 and EBT3 films. Calibrations of films for 4 MV x-rays were obtained for comparison using a radiotherapy accelerator at Mary Bird Perkins Cancer Center. RESULTS: The sensitivity (NOD per unit dose) of EBT film at 35 keV relative to that for 4-MV x-rays was 0.73 and 0.76 for doses 50 and 100 cGy, respectively. The sensitivity of EBT2 film at 25, 30, and 35 keV relative to that for 4-MV x-rays varied from 1.09-1.07, 1.23-1.17, and 1.27-1.19 for doses 50-200 cGy, respectively. For EBT3 film the relative sensitivity was within 3% of unity for all three monochromatic x-ray beams. CONCLUSIONS: EBT and EBT2 film sensitivity showed strong energy dependence over an energy range of 25 keV-4 MV, although this dependence becomes weaker for larger doses. EBT3 film shows weak energy dependence, indicating that it would be a better dosimeter for kV x-ray beams where beam hardening effects can result in large changes in the effective energy.


Assuntos
Dosimetria Fotográfica/instrumentação , Síncrotrons , Raios X , Relação Dose-Resposta à Radiação , Desenho de Equipamento , Análise de Falha de Equipamento , Doses de Radiação , Reprodutibilidade dos Testes , Sensibilidade e Especificidade