RESUMO
Objectives: This study aimed to objectively evaluate the water-jet-functioned electrosurgical knife injection performances in a desktop experiment. Methods: Five types of water-jet-functioned electrosurgical knives, including two injection styles of sheath-type (A: DualKnife J, KD-655L; B: FlushKnife, DK2620-J-B20S; C: Splash M-Knife, DN-D2718B; D: ISSEN, SN1650-20) and tip-type (E: ORISE ProKnife, M00519361) were evaluated. These knives were compared with an injection needle (Control: SuperGrip 25G) as a control. The injection speed under constant pressure and the injection efficiency for each knife against prepared porcine stomach mucosa were evaluated. The additional clear gel injections using an injection needle were observed using an indigo blue-colored gel to evaluate the difference between the locations of water-jet holes. Results: Four types of knives, except for A, showed significantly higher water-jet speeds (A: 0.79 ± 0.03 g/20 s, B: 2.56 ± 0.05 g/20 s, C: 3.09 ± 0.06 g/20 s, D: 2.86 ± 0.05 g/20 s, and E: 1.79 ± 0.03 g/20 s) compared to that of the control (1.21 ± 0.03 g/20 s). Meanwhile, significantly higher efficacy of injection was found in the tip-type water-jet function knife, second to the injection needle (Control: 37.2% ± 35.5%, A: 20.9% ± 20.2%, B: 1.1% ± 2.2%, C: 6.2% ± 12.6%, D: 12.5% ± 15.6%, and E: 33.3% ± 32.2%). An additional injection experiment revealed that the injection with a piercing tip into the gel could achieve sufficient additional injection inside the stacked clear gel. Conclusions: The tip-type water-jet function electrosurgical knife is preferable for effective submucosal injection during endoscopic treatments.
RESUMO
Video 1Crush the large-sized stones directly using an electrohydraulic lithotripsy device.
RESUMO
Background and study aims Snare devices play an important role in treatment of intestinal polyps. However, there are no objective measurements for the characteristics of the various types of snare devices. Materials and methods Seven types of snare devices from four manufacturers were evaluated based on original measurements. The evaluated factors were stiffness, cutting quality, and change in force required for cutting depending on sheath shape. The latter two factors were evaluated by measuring the force required to cut 20â% gelatin cylinders, which simulated intestinal polyps. The cutting sharpness was evaluated by observing the sectional surface of cut gelatin cylinders using a stereomicroscope. The correlations between these measurements and characteristics of the snare devices were investigated. Results A strong positive correlation, with an R 2 value of 0.863, was shown between the force required to cut gelatin cylinders and loop wire diameter. Loop wire diameter also had a strong correlation, with an R 2 value of 0.7997, with the change in force required for cutting gelatin cylinders depending on sheath shape. No correlations were detected between loop stiffness and characteristics of snare devices. The edge-enhanced image revealed that the rougher surfaces of the gelatin cylinders were cut by snares with a thicker diameter. Conclusions Thinner loop wire may provide higher performance in cold snare polypectomy in an experimental model.
RESUMO
Background and study aims Ultrathin endoscopes are commonly used for surveillance esophagogastroduodenoscopy (EGD) to reduce discomfort associated with scope insertion. However, the flexibility of an ultrathin endoscope is a trade-off between reducing discomfort and lengthening examination time. Patients and methods The EG17-J10 (EG17) is a novel ultrathin endoscope characterized by its tapering body stiffness; however, the flexibility of its tip is comparable to that of the traditional ultrathin endoscope EG16-K10 (EG16). We compared EGD examination time between EG17 and EG16.âA total of 319 examinees who underwent EGD from November 2019 to January 2020 at the Chiba-Nishi General Hospital were enrolled. Six examinees were excluded due to past history of surgical resection of the upper gastrointestinal tract or too much food residues; 313 examinees (EG17, 209; EG16,104) were retrospectively analyzed. The examination time was divided into three periods: esophageal insertion time (ET), gastroduodenal insertion time (GDT), and surveillance time of the stomach (ST). The total amount of ET, GDT, and ST was defined as total examination time (TT). Results TT of EGD using EG17 was significantly shorter compared to EGD using EG16 (222.7â±â68.9 vs. 245.7â±â78.5 seconds) ( P â=â0.004). Among the three periods of examination time, ET (66.7â±â24.1 vs. 76.0â±â24.1 seconds) ( P â=â0.001) and GDT (47.9â±â17.4 vs. 55.2â±â35.2 seconds) ( P â=â0.007) of EGD using EG17 were significantly shorter compared to EGD using EG16, except for ST (108.1â±â51.5.1 vs. 114.5â±â50.1 seconds) ( P â=â0.148). Conclusion An ultrathin endoscope with tapering body stiffness can shorten EGD examination time, mainly due to the shortening of insertion time.
RESUMO
A 54-year-old man with diarrhea and hematochezia for 2 months was referred to our department. A total colonoscopy revealed amoebic colitis caused by Entamoeba histolytica. Concurrently, a submucosal tumor-like yellowish hemispheric polypoid lesion was incidentally detected in the rectum. We speculated that the lesion was a NET, which could be cured by endoscopic treatment. However, histopathological assessment of a biopsy specimen unexpectedly revealed a well- or moderately differentiated adenocarcinoma. After treatment of the amoebic colitis, endoscopic ultrasound revealed a low, hetero-echoic, 6-mm-diameter lesion mainly in the submucosal layer. We performed surgical resection because the invasion was estimated to be to the deeper submucosal layer. Histopathological assessment of the surgically resected specimen revealed a focal lesion of a well-differentiated adenocarcinoma in the granulation tissue of the submucosal layer. In cases accompanied by amoebic colitis, a tumor's initial gross type might change. Diagnostic endoscopic resection could be acceptable in such cases.
Assuntos
Adenocarcinoma , Disenteria Amebiana , Tumores Neuroendócrinos , Neoplasias Retais , Adenocarcinoma/diagnóstico , Adenocarcinoma/cirurgia , Humanos , Masculino , Pessoa de Meia-Idade , Mucosa , Tumores Neuroendócrinos/diagnóstico , Tumores Neuroendócrinos/cirurgia , Neoplasias Retais/diagnóstico , Neoplasias Retais/cirurgiaAssuntos
Ductos Biliares Intra-Hepáticos , Cálculos Biliares/cirurgia , Fístula Intestinal/cirurgia , Litotripsia/métodos , Cirurgia Endoscópica por Orifício Natural/métodos , Feminino , Cálculos Biliares/complicações , Cálculos Biliares/diagnóstico por imagem , Humanos , Fístula Intestinal/complicações , Fístula Intestinal/diagnóstico por imagem , Pessoa de Meia-IdadeRESUMO
OBJECTIVE: To evaluate the usefulness of a training program on endoscopic head and neck surveillance for beginner endoscopists. METHODS: This prospective multicenter study included 13 beginner endoscopists from 10 institutions who received training in systematic observation techniques and diagnostic criteria, and the training involved hands-on learning. Between May 2016 and February 2017, enrolled patients with current or previously diagnosed esophageal squamous cell carcinomas underwent head and neck surveillance using narrow band imaging (NBI) endoscopy, and histologically confirmed head and neck squamous cell carcinoma (HNSCC) detection rates, endoscopic image quality, and examination times were compared before (group A) and after (group B) the training program. Maximum possible score for the endoscopic images was 30 points. RESULTS: A total of 330 patients, comprising 181 in group A and 149 in group B, were enrolled. Three patients with HNSCC were detected in group A (1.7%) and in group B (2.0%; P = 1.000). Mean ± standard deviation (SD) examination times were 157 ± 71 s and 174 ± 109 s in groups A and B, respectively, (P = 0.073). Mean ± SD scores of the endoscopic images were 25.04 ± 5.47 points and 27.01 ± 4.35 points in groups A and B, respectively, (P < 0.001). CONCLUSION: The HNSCC detection rate based on the use of NBI on patients with ESCC did not improve after the training program for beginner endoscopists; however, endoscopic image quality improved significantly after the training program.