1.
Phys Med Biol ; 68(5)2023 03 02.
Artigo em Inglês | MEDLINE | ID: mdl-36763999

RESUMO

Objective.T1 mapping of the liver is time consuming and can be challenging due to respiratory motion. Here we present a prospective slice tracking approach, which utilizes an external ultra-wide band radar signal and allows for efficient T1 mapping during free-breathing.Approach.The fast radar signal is calibrated to an MR-based motion signal to create a motion model. This motion model provides motion estimates, which are used to carry out slice tracking for any subsequent clinical scan. This approach was evaluated in simulations, phantom experiments andin vivoscans.Main results.Radar-based slice tracking was implemented on an MR system with a total latency of 77 ms. Moving phantom experiments showed accurate motion prediction with an error of 0.12 mm in anterior-posterior and 0.81 mm in head-feet direction. The model error remained stable for up to two hours.In vivoexperiments showed visible image improvement with a motion model error three times smaller than with a respiratory bellow. For T1 mapping during free-breathing the proposed approach provided similar results compared to reference T1 mapping during a breathhold.Significance.The proposed radar-based approach achieves accurate slice tracking and enables efficient T1 mapping of the liver during free-breathing. This motion correction approach is independent from scanning parameters and could also be used for applications like MR guided radiotherapy or MR Elastography.


Assuntos
Imageamento por Ressonância Magnética , Radar , Estudos Prospectivos , Imageamento por Ressonância Magnética/métodos , Movimento (Física) , Fígado/diagnóstico por imagem , Respiração , Imagens de Fantasmas
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA