1.
Ann Allergy ; 71(2): 103-6, 1993 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-8346860

RESUMO

Metered dose inhalers are difficult for patients to use. A device that eliminates coordination and timing of actuation may simplify the use of metered dose inhalers. This trial compared (1) number of errors made and (2) specific errors made between the conventional press and breathe metered dose inhaler (MDI) and the novel breath actuated Autohaler inhalation device in 24 subjects. We studied the use of each device in 12 patients trained and experienced in using an MDI and in 12 volunteers who had never been exposed to any inhalation device. We observed that even experienced patients continue to have difficulty with the coordination and timing of metered dose inhalers. The volunteer group had equal difficulty with both devices but it appeared that it was easier for them to learn how to use the breath actuated device than the MDI.


Assuntos
Asma/tratamento farmacológico , Nebulizadores e Vaporizadores/normas , Administração por Inalação , Adulto , Feminino , Humanos , Masculino , Erros de Medicação , Pessoa de Meia-Idade , Terapia Respiratória/métodos
2.
J Asthma ; 30(5): 373-9, 1993.
Artigo em Inglês | MEDLINE | ID: mdl-8407737

RESUMO

The Autohaler actuator is a breath-actuated device designed by 3M Pharmaceuticals. The objective of this study was to measure and improve the effectiveness of the device's package insert instructions (PII). Using only the PII for guidance, 5 of 20 (25%) subjects failed to trigger the device. The PII were revised based on the subject's performance and suggestions. Using the revised PII, only 1 of 20 (5%) different subjects failed to trigger the device. These results indicate that relatively minor modifications to instructions can result in significant improvements in patient use. Eighty-five percent of the participants thought the device was easier to use than a metered-dose inhaler.


Assuntos
Asma/tratamento farmacológico , Nebulizadores e Vaporizadores , Rotulagem de Produtos , Adolescente , Adulto , Criança , Desenho de Equipamento , Estudos de Avaliação como Assunto , Humanos