Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 4 de 4
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Rev Sci Instrum ; 86(8): 083105, 2015 Aug.
Artigo em Inglês | MEDLINE | ID: mdl-26329172

RESUMO

The extension of the detectable elemental range with Total Reflection X-ray Fluorescence (TXRF) analysis is a challenging task. In this paper, it is demonstrated how a TXRF spectrometer is modified to analyze elements from carbon to uranium. Based on the existing design of a vacuum TXRF spectrometer with a 12 specimen sample changer, the following components were renewed: the silicon drift detector with 20 mm(2) active area and having a special ultra-thin polymer window allowing the detection of elements from carbon upwards. Two exchangeable X-ray sources guarantee the efficient excitation of both low and high Z elements. These X-ray sources were two light-weighted easily mountable 35 W air-cooled low-power tubes with Cr and Rh anodes, respectively. The air cooled tubes and the Peltier-cooled detector allowed to construct a transportable tabletop spectrometer with compact dimensions, as neither liquid nitrogen cooling for the detector nor a water cooling circuit and a bulky high voltage generator for the X-ray tubes are required. Due to the excellent background conditions as a result of the TXRF geometry, detection limits of 150 ng for C, 12 ng for F, and 3.3 ng for Na have been obtained using Cr excitation in vacuum. For Rh excitation, the detection limits of 90 pg could be achieved for Sr. Taking 10 to 20 µl of sample volume, extrapolated detection limits in the ng/g (ppb) range are resulting in terms of concentration.

2.
Anal Bioanal Chem ; 400(8): 2649-54, 2011 Jun.
Artigo em Inglês | MEDLINE | ID: mdl-21190102

RESUMO

Total reflection X-ray fluorescence analysis (TXRF) offers a nondestructive qualitative and quantitative analysis of trace elements. Due to its outstanding properties TXRF is widely used in the semiconductor industry for the analysis of silicon wafer surfaces and in the chemical analysis of liquid samples. Two problems occur in quantification: the large statistical uncertainty in wafer surface analysis and the validity of using an internal standard in chemical analysis. In general TXRF is known to allow for linear calibration. For small sample amounts (low nanogram (ng) region) the thin film approximation is valid neglecting absorption effects of the exciting and the detected radiation. For higher total amounts of samples deviations from the linear relation between fluorescence intensity and sample amount can be observed. This could be caused by the sample itself because inhomogeneities and different sample shapes can lead to differences of the emitted fluorescence intensities and high statistical errors. The aim of the study was to investigate the elemental distribution inside a sample. Single and multi-element samples were investigated with Synchrotron-radiation-induced micro X-ray Fluorescence Analysis (SR-µ-XRF) and with an optical microscope. It could be proven that the microscope images are all based on the investigated elements. This allows the determination of the sample shape and potential inhomogeneities using only light microscope images. For the multi-element samples, it was furthermore shown that the elemental distribution inside the samples is homogeneous. This justifies internal standard quantification.

3.
Anal Chem ; 82(18): 7744-51, 2010 Sep 15.
Artigo em Inglês | MEDLINE | ID: mdl-20718426

RESUMO

Selenium content of soils is an important issue due to the narrow range between the nutritious requirement and toxic effects upon Se exposure. However, its determination is challenging due to low concentrations within complex matrices that hamper the analysis in most spectroscopic techniques. In this study, we explored the possibilities of several analytical approaches combined with total reflection X-ray (TXRF) spectrometry for soil Se determinations. The direct analysis of a solid suspension using 20 mg of fine ground material (<50 µm) has a relatively high Se limit of detection (LOD) of 1 mg/kg (worldwide Se average in soils = 0.4 mg/kg) and is therefore only suitable for seleniferous soils. Several fast and simple analytical strategies were developed to decrease matrix effects and improve the LOD for Se determination in soil digests. On one hand, the application of a liquid-liquid extraction procedure using ethyl ether and the introduction of a Cr absorbent in the instrument configuration were carried out to avoid the associated problems on TXRF analysis of soil extracts due to the high Fe concentrations (∼700 mg/L). On the other hand, a dispersive liquid-liquid microextraction procedure (DLLME) before the TXRF analysis of the soil digest was also developed. The effects of various experimental parameters such as sample volume, effect of major elements present in the soil matrix (Fe), and Se concentration in the sample were investigated. The LOD using this analytical methodology (0.05 mg/kg of Se) was comparable to or lower than those obtained in previous works using other popular spectrometric techniques such as GFAAS, ICPMS, and AFS. The calculated Se concentration for JSAC-0411 ([Se] = 1.32 ± 0.27 mg/kg) using the combination of DLLME and TXRF ([Se] = 1.40 ± 0.23 mg/kg) was in agreement with the certified value.


Assuntos
Selênio/análise , Solo/química , Análise Espectral/métodos , Fracionamento Químico , Ferro/química , Selênio/isolamento & purificação , Suspensões
4.
Talanta ; 82(2): 821-7, 2010 Jul 15.
Artigo em Inglês | MEDLINE | ID: mdl-20602976

RESUMO

At present, there is a considerable interest in Hg monitoring in wastewater samples due to its widespread occurrence and the high toxicity of most of its compounds. Hg determination in water samples by means of total reflection X-ray fluorescence spectrometry (TXRF) entails some difficulties due to the high vapor pressure and low boiling point of this element that produce evaporation and loss of Hg from the surface of the reflector during the drying process, commonly used for sample preparation in TXRF analysis. The main goal of the present research was to develop a fast and simple chemical strategy to avoid Hg volatilization during the analysis of wastewater samples by TXRF spectrometry. Three different analytical procedures were tested for this purpose: (i) increasing the viscosity of the wastewater sample by adding a non-ionic surfactant (Triton X-114), (ii) Hg immobilization on the quartz reflectors using the extractant tri-isobutylphosphine (Cyanex 471X) and (iii) formation of a stable and non-volatile Hg complex into the wastewater sample. The best analytical strategy was found to be the formation of a Hg complex with thiourea (pH=10) before the deposition of 10 microL of sample on the reflector for following TXRF analysis. Analytical figures of merit such as linearity, limits of detection, accuracy and precision were carefully evaluated. Finally, the developed methodology was applied for the determination of Hg in different types of wastewater samples (industrial effluents, municipal effluents from conventional systems and municipal effluents from constructed wetlands).


Assuntos
Mercúrio/análise , Poluentes Químicos da Água/análise , Absorção , Limite de Detecção , Mercúrio/química , Octoxinol , Compostos Organometálicos/análise , Compostos Organometálicos/química , Polietilenoglicóis/química , Espectrometria por Raios X/métodos , Propriedades de Superfície
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA