Rev Sci Instrum ; 89(10): 10F111, 2018 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-30399723

RESUMO

The paper reports on the optimization process of the soft X-ray pulse height analyzer installed on the Wendelstein 7-X (W7-X) stellarator. It is a 3-channel system that records X-ray spectra in the range from 0.6 to 19.6 keV. X-ray spectra, with a temporal and spatial resolution of 100 ms and 2.5 cm (depending on selected slit sizes), respectively, are line integrated along a line-of-sight that crosses near to the plasma center. In the second W7-X operation phase with a carbon test divertor unit, light impurities, e.g., carbon and oxygen, were observed as well as mid- to high-Z elements, e.g., sulfur, chlorine, chromium, manganese, iron, and nickel. In addition, X-ray lines from several tracer elements have been observed after the laser blow-off injection of different impurities, e.g., silicon, titanium, and iron, and during discharges with prefill or a gas puff of neon or argon. These measurements were achieved by optimizing light absorber-foil selection, which defines the detected energy range, and remotely controlled pinhole size, which defines photon flux. The identification of X-ray lines was confirmed by other spectroscopic diagnostics, e.g., by the High-Efficiency XUV Overview Spectrometer, HEXOS, and high-resolution X-ray imaging spectrometer, HR-XIS.