1.
Anal Bioanal Chem ; 398(5): 1871-81, 2010 Nov.
Artigo em Inglês | MEDLINE | ID: mdl-20658281

RESUMO

With a view toward developing a rugged microflow cytometer, a sheath flow system was micromachined in hard plastic (polymethylmethacrylate) for analysis of particles and cells using optical detection. Six optical fibers were incorporated into the interrogation region of the chip, in which hydrodynamic focusing narrowed the core stream to ~35 µm × 40 µm. The use of a relatively large channel at the inlet as well as in the interrogation region (375 µm × 125 µm) successfully minimized the risk of clogging. The device could withstand pressures greater than 100 psi without leaking. Assays using both coded microparticles and cells were demonstrated using the microflow cytometer. Multiplexed immunoassays detected nine different bacteria and toxins using a single mixture of coded microspheres. A549 cancer cells processed with locked nucleic acid probes were evaluated using fluorescence in situ hybridization.


Assuntos
Bioensaio/instrumentação , Citometria de Fluxo/instrumentação , Citometria de Fluxo/métodos , Bioensaio/métodos , Linhagem Celular Tumoral , Humanos , Imunoensaio , Microesferas , Polimetil Metacrilato/química