1.
Philos Trans A Math Phys Eng Sci ; 374(2064): 20150045, 2016 Mar 28.
Artigo em Inglês | MEDLINE | ID: mdl-26903100

RESUMO

We describe the design, construction, calibration and use of a near-infrared thermodynamic radiation thermometer to measure blackbodies from 400 K to 1300 K. The motivation for this work is the pending redefinition of the kelvin and the need for direct, thermodynamic temperature measurements of the fixed-point blackbodies presently used in the realization of the temperature scale. The challenges of accurately measuring Planck radiances which vary greatly in radiance level and spectral shape are discussed. Methods to characterize the components used in the radiation thermometer design are described. The use of this radiation thermometer as a relative primary thermometer and the resulting residuals are shown. We describe radiometric calibration procedures for using the radiation thermometer as an absolute primary thermometer. Preliminary data showing the initial radiometric calibration steps are discussed.

2.
J Res Natl Inst Stand Technol ; 118: 396-402, 2013.
Artigo em Inglês | MEDLINE | ID: mdl-26401440

RESUMO

We report a measurement of lunar spectral irradiance with an uncertainty below 1 % from 420 nm to 1000 nm. This measurement uncertainty meets the stability requirement for many climate data records derived from satellite images, including those for vegetation, aerosols, and snow and ice albedo. It therefore opens the possibility of using the Moon as a calibration standard to bridge gaps in satellite coverage and validate atmospheric retrieval algorithms. Our measurement technique also yields detailed information about the atmosphere at the measurement site, suggesting that lunar observations are a possible solution for aerosol monitoring during the polar winter and can provide nighttime measurements to complement aerosol data collected with sun photometers. Our measurement, made with a novel apparatus, is an order of magnitude more accurate than the previous state-of-the-art and has continuous spectral coverage, removing the need to interpolate between filter passbands.

3.
Appl Opt ; 51(16): 3631-41, 2012 Jun 01.
Artigo em Inglês | MEDLINE | ID: mdl-22695603

RESUMO

An algorithm is presented that corrects a multichannel fiber-coupled spectrograph for stray or scattered light within the system. The efficacy of the algorithm is evaluated based on a series of validation measurements of sources with different spectral distributions. This is the first application of a scattered-light correction algorithm to a multichannel hyperspectral spectrograph. The algorithm, based on characterization measurements using a tunable laser system, can be extended to correct for finite point-spread response in imaging systems.

4.
J Res Natl Inst Stand Technol ; 116(2): 621-46, 2011.
Artigo em Inglês | MEDLINE | ID: mdl-26989588

RESUMO

The pre-launch characterization and calibration of remote sensing instruments should be planned and carried out in conjunction with their design and development to meet the mission requirements. The onboard calibrators such as blackbodies and the sensors such as spectral radiometers should be characterized and calibrated using SI traceable standards. In the case of earth remote sensing, this allows inter-comparison and intercalibration of different sensors in space to create global time series of climate records of high accuracy where some inevitable data gaps can be easily bridged. The recommended best practice guidelines for this pre-launch effort is presented based on experience gained at National Institute of Standards and Technology (NIST), National Aeronautics and Space Administration (NASA) and National Oceanic and Atmospheric Administration (NOAA) programs over the past two decades. The currently available radiometric standards and calibration facilities at NIST serving the remote sensing community are described. Examples of best practice calibrations and intercomparisons to build SI (international System of Units) traceable uncertainty budget in the instrumentation used for preflight satellite sensor calibration and validation are presented.

5.
Appl Opt ; 49(10): 1859-64, 2010 Apr 01.
Artigo em Inglês | MEDLINE | ID: mdl-20357870

RESUMO

Results are presented for modeling of the shape of the internal quantum efficiency (IQE) versus wavelength for silicon photodiodes in the 400 nm to 900 nm wavelength range. The IQE data are based on measurements of the external quantum efficiencies of three transmission optical trap detectors using an extensive set of laser wavelengths, along with the transmittance of the traps. We find that a simplified version of a previously reported IQE model fits the data with an accuracy of better than 0.01%. These results provide an important validation of the National Institute of Standards and Technology (NIST) spectral radiant power responsivity scale disseminated through the NIST Spectral Comparator Facility, as well as those scales disseminated by other National Metrology Institutes who have employed the same model.

6.
J Res Natl Inst Stand Technol ; 105(5): 689-700, 2000.
Artigo em Inglês | MEDLINE | ID: mdl-27551631

RESUMO

A cryogenic radiometer-based system was constructed at the National Institute of Standards and Technology for absolute radiometric measurements to improve detector spectral power responsivity scales in the wavelength range from 900 nm to 1800 nm. In addition to the liquid-helium-cooled cryogenic radiometer, the system consists of a 100 W quartz-tungsten-halogen lamp light source and a 1 m single-grating monochromator for wavelength selection. The system was characterized and the uncertainty in spectral power responsivity measurements evaluated. A variety of photodetectors, including indium gallium arsenide photodiodes (InGaAs), germanium (Ge) photodiodes, and pyroelectric detectors, were subsequently calibrated. Over most of the spectral range, the spectral power responsivity of the photodetectors can be measured with a combined relative standard uncertainty of 0.4 % or less. This is more than a factor of two smaller than our previous capabilities, and represents a significant improvement in the near infrared (NIR) spectral power responsivity scale maintained at NIST. We discuss the characterization of the monochromator-based system and present results of photodetector spectral power responsivity calibrations.

7.
Appl Opt ; 38(1): 18-28, 1999 Jan 01.
Artigo em Inglês | MEDLINE | ID: mdl-18305582

RESUMO

The combination of a cryogenic radiometer and synchrotron radiation enables detector scale realization in spectral regions that are otherwise difficult to access. Cryogenic radiometry is the most accurate primary detector-based standard available to date, and synchrotron radiation gives a unique broadband and continuous spectrum that extends from x ray to far IR. We describe a new cryogenic radiometer-based UV radiometry facility at the Synchrotron Ultraviolet Radiation Facility II at the National Institute of Standards and Technology. The facility is designed to perform a variety of detector and optical materials characterizations. The facility combines a high-throughput, normal incidence monochromator with an absolute cryogenic radiometer optimized for UV measurements to provide absolute radiometric measurements in the spectral range from 125 nm to approximately 320 nm. We discuss results on photodetector characterizations, including absolute spectroradiometric calibration, spatial responsivity mapping, spectroreflectance, and internal quantum efficiency. In addition, such characterizations are used to study UV radiation damage in photodetectors that can shed light on the mechanism of the damage process. Examples are also given for UV optical materials characterization.

8.
Anal Chem ; 69(21): 4331-8, 1997 Nov 01.
Artigo em Inglês | MEDLINE | ID: mdl-21639167

RESUMO

Biotin and biotinylated self-assembled monolayers (SAMs) on gold have been investigated using time-of-flight secondary ion mass spectrometry, direct laser desorption, laser desorption with 193 nm photoionization of ion- and laser-desorbed species, and laser desorption with vacuum ultraviolet (VUV, 118 nm) photoionization. Our results indicate that direct laser desorption and laser desorption combined with 193 nm multiphoton ionization can detect a chromophoric molecule like biotin that is covalently bound to a SAM. However, secondary ion mass spectra were dominated by fragmentation, and ion desorption/193 nm photoionization detected no species related to biotin. The dominant features of the laser desorption/VUV mass spectra were neat and Au-complexed dimers of intact and fragmented biotinylated SAM molecules. Multiphoton and single-photon ionization of laser-desorbed neutrals from biotinylated SAMs both led to the production of ions useful for chemical analysis of the monolayer. Multiphoton ionization with ultraviolet radiation was experimentally less challenging but required a chromophore for ionization and resulted in significant fragmentation of the adsorbate. Single-photon ionization with VUV radiation was experimentally more challenging but did not require a chromophore and led to less fragmentation. X-ray photoelectron spectra indicated that the biotinylated SAM formed a disordered, 40-60 Å thick monolayer on Au. Additionally, projection photolithography with a Schwarzschild microscope was used to pattern the biotinylated SAM surface and laser desorption/photoionization was used to detect biotinylated adsorbates from the ∼10 µm sized pattern.

9.
Anal Chem ; 68(5): 713-9, 1996 Mar 01.
Artigo em Inglês | MEDLINE | ID: mdl-21619164

RESUMO

A microscopic laser desorption/postionization Fourier transform mass spectrometer (LD/FTMS) is described. The lateral resolution can be <1 µm with the inherent FTMS high mass resolution intact. Laser postionization allows a certain selectivity and an increase in sensitivity. This capability should allow materials characterization in a wide variety of cases. We demonstrate microscopic desorption and postionization for atoms and molecules.

10.
Appl Opt ; 32(6): 857-66, 1993 Feb 20.
Artigo em Inglês | MEDLINE | ID: mdl-20802761

RESUMO

A powerful molecular surface analysis technique for the analysis of complex materials, such as polymer/additive systems, consists of laser desorption of surface molecules and subsequent ionization of these gas-phase molecules with resonant or nonresonant laser ionization. These molecular ions are subsequently detected by Fourier-transform mass spectrometry or time-of-flight mass spectrometry. We show that different wavelengths for the postionization step permit selectivity that provides important additional information on the chemical makeup of these complex materials. Near-UV wavelengthsselectively ionize aromatic polymer additives, far-UV wavelengths photoionize other nonaromatic species; and vacuum-UV wavelengths provide access to all the desorbed species. In addition to these applied results, we study many fundamental issues of laser desorption, such as desorption thresholds, velocity distributions, postionization wavelength selectivity, etc. The Fourier-transform mass spectrometer and time-of-flight mass spectrometer are discussed in detail.