Clin Biochem ; 42(15): 1582-4, 2009 Oct.
Artigo em Inglês | MEDLINE | ID: mdl-19615990

RESUMO

OBJECTIVES: To compare HPLC methods with short and long elution times for HbA1c measurement in blood. METHODS: Comparison between G7-Tosoh (1.2 min); Bio-Rad-Variant-II-Turbo (1.3 min) and Arkray-HA-8160 (2.9 min). RESULTS: Passing-Bablok regression equations were: Y=0.17+0.96X; Y=-0.39+1.01X; Y=-0.40+1.0X and the means of the differences using Bland-Altman Plot were 0.02; -0.34; 0.32 for HA-8160/G7-Tosoh, HA-8160/Variant-II-Turbo and G7-Tosoh/Variant-II-Turbo, respectively. CONCLUSIONS: Faster elution methods had no problems on reproducibility of results obtained by slower elution methods.