Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros











Base de dados
Intervalo de ano de publicação
1.
Doc Ophthalmol ; 138(3): 169-179, 2019 06.
Artigo em Inglês | MEDLINE | ID: mdl-30840173

RESUMO

PURPOSE: Visual evoked potentials (VEPs) are used to assess visual function in preclinical models of neurodegenerative diseases. VEP recording with epidural screw electrodes is a common method to study visual function in rodents, despite being an invasive procedure that can damage the tissue under the skull. The present study was performed to test a semi-invasive (epicranial) and a non-invasive (epidermal) VEP recording technique, comparing them with the classic epidural acquisition method. METHODS: Flash VEPs were recorded from C57BL/6 mice on three separate days within 2 weeks. Waveforms, latencies and amplitudes of the components were compared between the three different methods, utilizing coefficient of repeatability, coefficient of variation and intersession standard deviation to evaluate reproducibility. RESULTS: While epidural electrodes succeeded in recording two negative peaks (N1 and N2), epicranial and epidermal electrodes recorded a single peak (N1). Statistical indexes showed a comparable reproducibility between the three techniques, with a greater stability of N1 latency recorded through epicranial electrodes. Moreover, N1 amplitudes recorded with the new less-invasive methods were more reproducible compared to the invasive gold-standard technique. CONCLUSIONS: These results demonstrate the reliability of semi- and non-invasive VEP recordings, which can be useful to evaluate murine models of neurological diseases.


Assuntos
Potenciais Evocados Visuais/fisiologia , Vias Visuais/fisiologia , Animais , Eletrodos , Eletrorretinografia , Epiderme , Feminino , Camundongos , Camundongos Endogâmicos C57BL , Estimulação Luminosa/métodos , Reprodutibilidade dos Testes
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA