1.
Clin Chem ; 35(12): 2333-5, 1989 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-2686863

RESUMO

An automated enzyme immunoassay for human lutropin for use with the Abbott IMx analyzer is described. The assay provides results in approximately 40 min with a sensitivity of 0.25 int. units of LH per liter for up to 23 serum or plasma samples. Cross-reactivity with follitropin (2000 int. units/L) and thyrotropin (2 int. units/L) was negligible; it was 0.016% with human choriogonadotropin (1 X 10(6) int. units/L). There was no interference by high concentrations of bilirubin (0.5 g/L), hemoglobin (7.50 g/L), or triglycerides (13.5 g/L). Intra-, inter-, and total assay CVs were less than or equal to 3.75%, less than or equal to 7.1%, and less than or equal to 7.94%, respectively. Values obtained with the IMx correlated well (r = 0.98, n = 194) with values obtained with Diagnostic Products' LH Double Antibody RIA, and Serono's LH MAIAclone assay. This assay should be useful for small to medium-size laboratories involved in the clinical diagnosis of reproductive pathology.


Assuntos
Hormônio Luteinizante/sangue , Amenorreia/sangue , Autoanálise/instrumentação , Reações Cruzadas , Feminino , Humanos , Técnicas Imunoenzimáticas , Masculino , Menopausa , Ciclo Menstrual , Puberdade , Kit de Reagentes para Diagnóstico , Estatística como Assunto
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA