1.
Appl Biochem Biotechnol ; 186(4): 1061-1073, 2018 Dec.
Artigo em Inglês | MEDLINE | ID: mdl-29862444

RESUMO

A cantilever nanobiosensor functionalized with vegetable source of peroxidase was developed as an innovative way for glyphosate herbicide detection over a wide concentration range (0.01 to 10 mg L-1) using atomic force microscopy (AFM) technique. The extract obtained from zucchini (Cucurbita pepo source of peroxidase), with high enzymatic activity and stability has been used as bio-recognition element to develop a nanobiosensor. The polarization-modulated reflection absorption infrared spectroscopy (PM-RAIRS) demonstrated the deposition of enzyme on cantilever surface using self-assembled monolayers (SAM) by the presence of the amide I and II bands. The detection mechanism of glyphosate was based on the changes in surface tension caused by the analyte adsorption, resulting in a conformational change in the enzyme structure. In this way, the results of nanobiosensor demonstrate the potential of the sensing device for detecting glyphosate with a detection limit of 0.028 mg L-1.


Assuntos
Técnicas Biossensoriais/métodos , Cucurbita/enzimologia , Glicina/análogos & derivados , Peroxidase/química , Proteínas de Plantas/química , Glicina/análise , Glifosato