Lab Med ; 48(1): 10-17, 2017 Feb.
Artigo em Inglês | MEDLINE | ID: mdl-27593171

RESUMO

OBJECTIVE: Validation of the LT-SYS quantitative in vitro copper assay on a Roche Cobas 8000 c502 analyzer and comparison with a BIOMED assay on a Roche Cobas Mira analyzer. METHODS: Imprecision and bias were quantified at different concentration levels (serum and plasma) over a 20-day period. Linearity was assessed covering a range from 4.08 µmol/L to 33.8 µmol/L. Limit of blank (LoB) and limit of detection (LoD) were established based on a total of 120 blank and low-level samples. The method comparison was based on 58 plasma samples. RESULTS: Within-run imprecision ranged from 0.7% to 1.2% and within-laboratory imprecision from 1.4% to 3.3%. Relative bias for the 2 serum pools with known target values was less than 2.5%. The assay did not deviate from linearity over the tested measuring range. LoB and LoD were 0.12 µmol/L and 0.23 µmol/L, respectively. The method comparison revealed an average deviation of 11.5% (2.016 µmol/L), and the linear regression fit was y = 1.464 + 0.795x. CONCLUSIONS: The LT-SYS copper assay characterized in this study showed a fully acceptable performance with good degrees of imprecision and bias, no deviation from linearity in the relevant measuring rangem, and very low LoB and LoD.


Assuntos
Testes de Química Clínica/métodos , Testes de Química Clínica/normas , Cobre/sangue , Adulto , Testes de Química Clínica/instrumentação , Feminino , Humanos , Limite de Detecção , Modelos Lineares , Masculino , Padrões de Referência , Reprodutibilidade dos Testes
SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...