Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 1 de 1
Filtrar
Mais filtros










Base de dados
Intervalo de ano de publicação
1.
Artigo em Inglês | MEDLINE | ID: mdl-18263138

RESUMO

The operation of an acoustic microscope having a resolution of 15 mm has been demonstrated. It uses as a coupling medium superfluid (4 )He colder than 0.9 K and pressurized to greater than 20 bar. The microscope has been used to image objects that show little or no contrast on a scanning electron microscope. In addition, the acoustic microscope has been used to study the properties of sound propagation in the coupling fluid. At low acoustic intensities, the coupling fluid has very low acoustic attenuation at the microscope's operating frequency (15.3 GHz), but near the focal point the acoustic intensity can be high enough that the helium behaves with extreme nonlinearity. In fact, this medium is capable of entering new regimes of nonlinear interaction. Plots of the received signal versus input power display a nearly complete source depletion at certain input power levels and a reconversion to the source frequency at higher power levels. Arguments that the process underlying this nonlinear behavior is harmonic generation are presented.

SELEÇÃO DE REFERÊNCIAS
DETALHE DA PESQUISA
...