RESUMO
OBJECTIVES: Placement of a proximal humerus locking plate through a percutaneous transdeltoid approach bears the advantages of a minimally invasive approach but may compromise the anterior branches of the axillary nerve. This anatomic study aimed to develop a risk profile for 6 types of modern proximal humerus locking plates as to their interference with the axillary nerve. MATERIALS AND METHODS: In this study six different implants (Arthrex®, DePuy®, Königsee®, Smith & Nephew®, Stryker® and Synthes®) were placed on the intact proximal humerus of 33 embalmed cadaveric upper extremities and the relative positioning between the axillary nerve and the screw holes was determined. RESULTS: All locking plates displayed an area of risk which concerned 3 out of 7 (Arthrex®), 4 out of 10 (DePuy®), 2 out of 9 (Königsee®), 3 out of 11 (Smith & Nephew®), 3 out of 11 (Stryker®) and 6 out of 12 (Synthes®) screw holes of the plate. CONCLUSIONS: Using the anterolateral percutaneous deltoid splitting approach the relative position of the axillary nerve to the holes of a specific implant is of relevance for avoidance of iatrogenic lesions to the nerve.